MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.plabm.2021.e00252

Медицинская библиотека · Practical laboratory medicine · 2021

The "EPiQ"-Study (Evaluation of preanalytical quality): S-Monovette® in manual aspiration mode drastically reduces hemolytic samples in head-to-head study.

Millius Laura, Riedo Erwin, Caron Thierry, Belissent Juliette, Fellay Benoît, Ribordy Vincent, Magnin Jean-Luc

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Hemolytic blood samples are the number one cause for specimen rejection at emergency departments. Triggered by unsuitable blood sampling material or incorrect handling and a related strong vacuum force, hemolytic samples often must be retaken. The objective of this study was to assess whether correct manual aspiration using S-Monovette® could reduce the number of hemolytic samples.

METHODS

Between January and April 2019, a head-to-head study was conducted. Whereas in the first eight weeks, all specimens were collected using Vacutainer®, in the second eight weeks, blood was taken using S-Monovette® in aspiration mode. Specimens were categorized into five classes (0-30, 31-50, 51-75, 76-100, and 101+ mg/dl of cell-free hemoglobin) and for the statistical analyses, all samples exceeding 30 mg/dl were classified as hemolytic.

RESULTS

Data were collected on 4794 blood specimens (Vacutainer®: 2634 samples, S-Monovette®: 2160 samples). While the percentage of non-hemolytic samples (HI of 0-30 mg/dl) was substantially higher for specimens drawn by S-Monovette® (95.7 %) than Vacutainer® (83.0 %), the opposite was true for all HI categories above 30 mg/dl. Importantly, the reduction of hemolytic samples took place immediately following the imposition of S-Monovette® and remained stable at a low level until the end of the study.

CONCLUSIONS

Based on our results, we conclude that switching to S-Monovette® in manual aspiration mode in the blood sampling process could be highly beneficial, not only from a financial point of view, but also with regards to reducing unnecessary tasks and stress for nursing staff and improving patient outcome overall.

Для цитирования: Millius Laura, Riedo Erwin, Caron Thierry, Belissent Juliette, Fellay Benoît, Ribordy Vincent, Magnin Jean-Luc. The "EPiQ"-Study (Evaluation of preanalytical quality): S-Monovette® in manual aspiration mode drastically reduces hemolytic samples in head-to-head study.. Practical laboratory medicine. 2021. DOI: 10.1016/j.plabm.2021.e00252.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.