DOI 10.5152/turkarchpediatr.2020.20183
Safety and feasibility of a factory-calibrated continuous glucose monitoring system in term and near-term infants at risk of hypoglycemia.
Аннотация
OBJECTIVE
Hypoglycemia increases the risk of adverse neurological outcomes in neonates. Adequate glucose monitoring requires repetitive and painful blood sampling. We aimed to evaluate the feasibility and accuracy of a continuous glucose monitoring system (CGMS) using factory-calibrated sensors to improve glucose monitoring and decrease the frequency of blood samples in neonates.
MATERIAL AND
METHODS
A methodological study was conducted to investigate a correlation of CGMS values with blood glucose measurements.
RESULTS
Factory-calibrated CGMS sensors were placed on 21 infants at risk of hypoglycemia after delivery. CGMS values were compared with blood glucose concentrations. Thirty-seven pairs of CGMS and blood glucose values were obtained. There was a good correlation between CGMS and blood glucose values (R=0.67, p<0.01) with a mean difference (2 standard deviations) of 9.78 (-24.68 to 44.25) mg/dL. The mean differences at <3 hours and ≥3 hours after sensor placement were 17.35 (-4.54 to 39.21) mg/dL and 0.88 (-37.62 to 39.38) mg/dL, respectively. CGMS values were significantly higher than blood glucose concentration at <3 hours after sensor placement (p<0.01), whereas no significant differences in glucose values were observed between the CGMS and blood glucose values at ≥3 hours after sensor placement (p=0.852).
CONCLUSION
The factory-calibrated CGMS was a safe and feasible modality for glucose monitoring. However, it has a tendency to overestimate the blood glucose concentrations. Therefore, this system should be used cautiously for neonates at risk of hypoglycemia, especially within 3 hours after sensor placement.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.