OBJECTIVE

Peripherally inserted central catheters (PICCs) are a mainstay of nonpermanent vascular access devices. In this study, we assessed patients displaying anaphylaxis or anaphylactoid reactions to the PowerPICC SOLO and Groshong PICC (Bard Access Systems) using the Sherlock tip locating system (TLS).

METHODS

Patients from 2 tertiary-care hospitals were systematically monitored over 4 years for adverse events following the insertion of a PICC using the Sherlock TLS. Insertion data were also collected using the BioFlo PICC (Angiodynamics)from a third hospital site and from The Ottawa Hospital over 4 years as an additional comparator. Three definitions of anaphylaxis and anaphylactoid reactions were utilized, and the Cohen κ was used to assess interrater agreement. Analysis of reactions among the patient cohorts was performed using the χ2 test with Yates correction or the Fisher exact test as appropriate.

RESULTS

Among 8,257 insertions using the TLS PICCs, 37 potential reactions (0.45%) were recorded. Using specific definitions for anaphylaxis or anaphylactoid reactions, 54.1%-91.9% met criteria. Comparator populations using data from Calgary (n = 491) and Ottawa (n = 7,889) using the BioFlo PICC insertion found no reactions. Anaphylactic or anaphylactoid reactions were significantly associated with the PowerPICC SOLO and Groshong PICC with the TLS compared to the BioFlo PICC (P < .0001) across all definitions. The largest subset of patients experiencing adverse reactions had cystic fibrosis (CF) (n = 4, 10.8%).

CONCLUSION

Our study results demonstrate significant adverse events associated with the PowerPICC SOLO and Groshong PICC using the Sherlock TLS inserted across a range of patient populations. The incidence rate of anaphylaxis or anaphylactoid reactions in the CF population at our center is significantly higher than in non-CF patients (P < .001).

Для цитирования: Thornton Christina S, Dumanski Jody, Margherit Cherylanne, Vaz-Gonsalves Sandra, McDiarmid Sheryl, Parkins Michael D, Conly John M. Anaphylaxis and anaphylactoid reactions associated with the insertion of peripherally inserted central catheters: A multiyear comparative retrospective cohort study.. Infection control and hospital epidemiology. 2019. DOI: 10.1017/ice.2019.237.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.