DOI 10.1016/j.arth.2019.03.027
Multicenter Evaluation of a Modular Dual Mobility Construct for Revision Total Hip Arthroplasty.
Аннотация
INTRODUCTION
Modular Dual Mobility (MDM) constructs in total hip arthroplasty (THA) offer increased hip stability compared with constrained liners, without compromising hip range of motion. The purpose of this study was to evaluate outcomes of revision THA using MDM.
METHODS
The study was a multiinstitutional retrospective cohort study of 315 hips that underwent revision THA using MDM between 2011 and 2017. Clinical outcomes and reasons for failure were collected.
RESULTS
Three hundred fifteen patients met 1-year minimum follow-up (mean 3.3 years). Nine hips had instability postoperatively (2.9%), and 30 hips required reoperation (9.5%). Seven had recurrent instability (6.5%) which was associated with liner-only exchange (P = .021) and liner outer diameter of ≤ 38 mm (P = .016).
CONCLUSION
Revision THA with MDM provided a low rate of instability and reoperation in a revision cohort. Recurrent instability following use of MDM in revision THA was associated with retention of the acetabular component and polyethylene outer diameter ≤ 38 mm.
LEVEL OF EVIDENCE: Therapeutic Level III.