MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.30697

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2023

Cerebral embolic protection during transcatheter aortic valve replacement: Insights from a consecutive series with the Sentinel cerebral protection device.

Wolfrum Mathias, Moccetti Federico, Loretz Lucca, Bossard Matthias, Attiger Adrian, Cuculi Florim, Toggweiler Stefan

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Growing interest in neuroprotection in transcatheter aortic valve replacement (TAVR) has catalyzed the development of cerebral protection systems (CPS).

OBJECTIVES

Report insights from consecutive real-world patients undergoing TAVR with the Sentinel-CPS.

METHODS

Patients with severe aortic stenosis undergoing TAVR from April 2019 to May 2022 were enrolled in a prospective registry. The reason for unsuccessful Sentinel-CPS deployment and the amount of debris captured by the filters were prospectively recorded.

RESULTS

The Sentinal CPS was successfully deployed in 330 patients (85%, Group 1). Deployment was not attempted, unsuccessful or only partially successful in 59 patients (15%, Group 2), caused by anatomical factors such as tortuosity, heavy calcification or small dimensions of radial or brachial artery in 46, technical aspects such as puncture failure or dissection in 5 or use of right radial access for the pigtail in 6. Debris was captured in 98% of patients in Group 1. In 40%, the amount of debris was graded moderate or extensive. Predictors for moderate/extensive debris were moderate/severe aortic calcification (OR 1.50, CI 1.05-2.15, p = 0.03), pre- and postdilatation (OR 1.97, CI 1.02-3.79, p = 0.04 and OR 1.71, CI 1.01-2.89, p = 0.048). The risk of stroke was numerically lower in patients who underwent TAVR with the Sentinel CPS (2.1 vs. 5.1%, respectively, p = 0.15). There was no stroke during CPS deployment, but one patient had a stroke immediately after device retrieval.

CONCLUSIONS

The Sentinel-CPS was successfully deployed in 85% of patients. Predictors for moderate/extensive debris captured were moderate/severe aortic calcification, pre- and postdilatation.

Для цитирования: Wolfrum Mathias, Moccetti Federico, Loretz Lucca, Bossard Matthias, Attiger Adrian, Cuculi Florim, Toggweiler Stefan. Cerebral embolic protection during transcatheter aortic valve replacement: Insights from a consecutive series with the Sentinel cerebral protection device.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2023. DOI: 10.1002/ccd.30697.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.