Conduction system pacing in difficult cardiac anatomies: Systematic approach with the 3D electroanatomic mapping guide.
Аннотация
INTRODUCTION
Restoring physiological cardiac electrical activity in patients with conduction disease can be crucial for the survival and quality of life. Conduction system pacing (CSP) is a valuable option, although it is limited by technical challenges in difficult anatomies. 3D electroanatomical mapping (3D-EAM) can support CSP ensuring high electro-anatomical precision and low fluoroscopy.
OBJECTIVES
We evaluated the feasibility and effectiveness of a systematic 3D-EAM use to guide CSP in difficult anatomical scenarios (highly dilated atria, congenital cardiomyopathies, failed biventricular implants (BiV) and pacing-induced cardiomyopathy (PICM)).
METHODS
Forty-three consecutive patients (27 males, 75 ± 10 years old) with standard pacing indications and difficult anatomical scenarios were included. The right atrium, His cloud, and atrio-ventricular septum were reconstructed by 3D-EAM. The His bundle (HB) was the initial target, while left bundle branch area pacing (LBBAP) was aimed at in case of unsatisfactory parameters, sub-optimally paced QRS, or impossibility of reaching the HB.
RESULTS
CSP was successful in 37 (86%) patients (15 HBP; 22 LBBAP). Mean mapping, fluoroscopy, and procedural times were 18 ± 7 min, 7 ± 5 min, 98 ± 47 min, respectively. The mean pacing threshold, R wave sensing, and pacing impedance of CSP lead were 1.2 ± 0.5V@0.5ms, 11.4 ± 6.2 mV, 736 ± 306 Ω, respectively. Baseline and paced QRS were 139 ± 38 ms and 114 ± 23 ms, respectively. No procedural complications were observed.
CONCLUSIONS
3D-EAM allowed the accurate definition of the His cloud and high ventricular septum and effectively guided CSP. It facilitated CSP in complex anatomies, with a procedural success rate of 86%. The results were satisfactory and reproducible, with acceptable fluoroscopy and procedural times.