DOI 10.1016/s0009-8981(01)00809-9

Медицинская библиотека · Clinica chimica acta; international journal of clinical chemistry · 2002

Detection of DNA mutations associated with mitochondrial diseases by Agilent 2100 bioanalyzer.

Lu Ching-You, Tso Dan-Ju, Yang Tom, Jong Yuh-Jyh, Wei Yau-Huei

Клинический случайОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Molecular analysis of mitochondrial DNA (mtDNA) has provided a final diagnosis for many of the mitochondrial diseases. We evaluated the Agilent 2100 bioanalyzer (Agilent Technologies, Palo Alto, CA) to determine whether the system could replace the conventional restriction fragment length polymorphism (RFLP) analysis by the agarose gel electrophoresis for the detection of the mtDNA mutation.

METHODS

Three members of a family with MELAS syndrome and four members of a family with MERRF syndrome were recruited for this study. After PCR and restriction enzyme digestion, DNA fragments were separated on the Agilent 2100 bioanalyzer in conjunction with the DNA 500 and DNA 1000 Labchip kits and by electrophoresis on precast 3% agarose gels.

RESULTS

The data generated by the DNA 500 and DNA 1000 assays using the Agilent 2100 bioanalyzer showed a lower percentage error and a better reproducibility as compared to those obtained by the conventional method.

CONCLUSION

Based on the performance of the bioanalyzer, we suggest that this novel Labchip is adequate to replace the current RFLP analysis by the agarose gel electrophoresis for mtDNA mutation detection.

Для цитирования: Lu Ching-You, Tso Dan-Ju, Yang Tom, Jong Yuh-Jyh, Wei Yau-Huei. Detection of DNA mutations associated with mitochondrial diseases by Agilent 2100 bioanalyzer.. Clinica chimica acta; international journal of clinical chemistry. 2002. DOI: 10.1016/s0009-8981(01)00809-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.