DOI 10.1016/j.jcin.2018.06.018

Медицинская библиотека · JACC. Cardiovascular interventions · 2018

Significant Differences in Debris Captured by the Sentinel Dual-Filter Cerebral Embolic Protection During Transcatheter Aortic Valve Replacement Among Different Valve Types.

Seeger Julia, Virmani Renu, Romero Maria, Gonska Birgid, Rottbauer Wolfgang, Wöhrle Jochen

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to evaluate the debris captured by the Claret Sentinel cerebral embolic dual-filter protection device during transfemoral transcatheter aortic valve replacement (TAVR) with different valve types.

BACKGROUND

Risk for embolization of debris during TAVR may vary by TAVR device.

METHODS

The filters of 100 consecutive patients were collected and captured debris was analyzed by histopathology and histomorphometry. Three valve types were implanted: the balloon-expandable Edwards SAPIEN 3 (n = 42), the self-expandable Medtronic Evolut R (n = 35), and the mechanically implantable Boston Scientific Lotus (n = 23).

RESULTS

Among the 3 groups there was no difference in baseline data, including Society of Thoracic Surgeons score for mortality, calcification, or pre-dilation. The type of captured debris did not differ among the 3 valve types in the proximal or distal filter. With the balloon-expandable valve, there were significantly more patients with large debris measuring ≥1,000 μm. The number of particles in the proximal filter was significantly lower with the Lotus (89.8 ± 106.3) compared with the Evolut R (187.3 ± 176.9) and Edwards SAPIEN 3 (172.3 ± 133.5) valves (p = 0.035). Total tissue area in the proximal filter was significantly smaller for the Lotus compared with the other 2 valve types (7.1 ± 6.3, 20.1 ± 19.0, and 21.3 ± 15.1 mm2; p = 0.0014). In contrast, for the distal filter, there were no differences with respect to valve type for total tissue area, particle size, and number of particles.

CONCLUSIONS

A significant difference was observed in the size and number of captured tissue particles with the double-filter embolic protection device among different valve types in patients undergoing TAVR. The largest particles were observed in patients treated with a balloon-expandable valve.

Для цитирования: Seeger Julia, Virmani Renu, Romero Maria, Gonska Birgid, Rottbauer Wolfgang, Wöhrle Jochen. Significant Differences in Debris Captured by the Sentinel Dual-Filter Cerebral Embolic Protection During Transcatheter Aortic Valve Replacement Among Different Valve Types.. JACC. Cardiovascular interventions. 2018. DOI: 10.1016/j.jcin.2018.06.018.