DOI 10.1016/j.jtcvs.2018.05.042

Медицинская библиотека · The Journal of thoracic and cardiovascular surgery · 2018

An in vitro evaluation of turbulence after transcatheter aortic valve implantation.

Hatoum Hoda, Yousefi Atieh, Lilly Scott, Maureira Pablo, Crestanello Juan, Dasi Lakshmi P

Сравнительное исследованиеСтатьяПоддержка NIH: внешняя программаVideo-Audio MediaПеревод пока не подготовлен

Аннотация

BACKGROUND

This study aimed at assessment of post-transcatheter aortic valve (TAV) replacement hemodynamics and turbulence when a same-size SAPIEN 3 (Edwards Lifesciences Corp, Irvine, Calif) and Medtronic Evolut (Minneapolis, Minn) were implanted in a rigid aortic root with physiological dimensions and in a representative root with calcific leaflets obtained from patient computed tomography scans.

METHODS

TAV hemodynamics were studied by placing a SAPIEN 3 26-mm and an Evolut 26-mm in rigid aortic roots and representative root with calcific leaflets under physiological conditions. Hemodynamics were assessed using high-fidelity particle image velocimetry and high-speed imaging. Transvalvular pressure gradients (PGs), pinwheeling indices, and Reynolds shear stress (RSS) were calculated.

RESULTS

(1) PGs obtained with the Evolut and the SAPIEN 3 were comparable among the different models (10.5 ± 0.15 mm Hg vs 7.76 ± 0.083 mm Hg in the rigid model along with 13.9 ± 0.19 mm Hg vs 5.0 ± 0.09 mm Hg in representative root with calcific leaflets obtained from patient computed tomography scans respectively); (2) more pinwheeling was found in the SAPIEN 3 than the Evolut (0.231 ± 0.057 vs 0.201 ± 0.05 in the representative root with calcific leaflets and 0.366 ± 0.067 vs 0.122 ± 0.045 in the rigid model); (3) higher rates of RSS were found in the Evolut (161.27 ± 3.45 vs 122.84 ± 1.76 Pa in representative root with calcific leaflets and 337.22 ± 7.05 vs 157.91 ± 1.80 Pa in rigid models). More lateral fluctuations were found in representative root with calcific leaflets.

CONCLUSIONS

(1) Comparable PGs were found among the TAVs in different models; (2) pinwheeling indices were found to be different between both TAVs; (3) turbulence patterns among both TAVs translated according to RSS were different. Rigid aortic models yield more conservative estimates of turbulence; (4) both TAVs exhibit peak maximal RSS that exceeds platelet activation 100 Pa threshold limit.

Для цитирования: Hatoum Hoda, Yousefi Atieh, Lilly Scott, Maureira Pablo, Crestanello Juan, Dasi Lakshmi P. An in vitro evaluation of turbulence after transcatheter aortic valve implantation.. The Journal of thoracic and cardiovascular surgery. 2018. DOI: 10.1016/j.jtcvs.2018.05.042.

Предметные рубрики MeSH