Usefulness of contrast injection during balloon aortic valvuloplasty before transcatheter aortic valve replacement: a pilot study.

Barbanti Marco, Sgroi Carmelo, Immè Sebastiano, Aruta Patrizia, Deste Wanda, Gulino Simona, Cannata Stefano, Giarratana Alessandra, Bottari Vera, Giannazzo Daniela, Garretto Valeria, Patanè Martina, Benvenuto Emanuele, Capodanno Davide, Tamburino Corrado

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

AIMS: To describe a technique of simultaneous aortography and balloon aortic valvuloplasty (BAV) before transcatheter aortic valve replacement (TAVR), and to show how this technique affected TAVR prosthesis selection and procedural outcomes.

METHODS AND

RESULTS

One hundred and eleven patients underwent simultaneous contrast injection during valvuloplasty pre-TAVR to confirm the indication for prosthesis size provided by non-invasive imaging studies. A successful injection was achieved in 95 patients (85.5%). No events occurred during simultaneous BAV and contrast injection. In 12 (10.8%) patients the prosthesis size implanted was different from the recommendations provided by the non-invasive imaging examinations. In nine of these cases (75.0%) it was decided to implant a larger prosthesis than that originally suggested, in the remaining three cases (25.0%) a smaller valve was implanted. Device success in this particular subset of patients was 100%. Overall device success was 92.8%. Post-procedural moderate paravalvular regurgitation was reported in 5.4% of patients.

CONCLUSIONS

In patients with severe aortic valve stenosis, a technique of simultaneous aortography and balloon valvuloplasty as an adjunct to non-invasive imaging modalities for transcatheter prosthesis selection is feasible, and leads to a change in TAVR strategy in a modest number of patients. Larger studies are necessary to confirm these findings, and to assess whether this method is capable of enhancing the safety of the TAVR procedure.

Для цитирования: Barbanti Marco, Sgroi Carmelo, Immè Sebastiano, Aruta Patrizia, Deste Wanda, Gulino Simona, Cannata Stefano, Giarratana Alessandra, Bottari Vera, Giannazzo Daniela, Garretto Valeria, Patanè Martina, Benvenuto Emanuele, Capodanno Davide, Tamburino Corrado. Usefulness of contrast injection during balloon aortic valvuloplasty before transcatheter aortic valve replacement: a pilot study.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2014. DOI: 10.4244/eijv10i2a39.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.