m+Medelevator.RU
← Поиск публикаций
ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2013

Rotary pumps and diminished pulsatility: do we need a pulse?

СтатьяОбзорНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Soucy Kevin G, Koenig Steven C, Giridharan Guruprasad A, Sobieski Michael A, Slaughter Mark S

Аннотация

Ventricular assist devices (VADs) have been successfully used as a bridge to heart transplant and destination therapy (DT) for congestive heart failure (HF) patients. Recently, continuous flow VAD (CVAD) has emerged as an attractive clinical option for long-term mechanical support of HF patients, with bridge-to-transplant outcomes comparable with pulsatile flow VAD (PVAD). Continuous flow VADs are smaller, more reliable, and less complex than the first-generation PVAD. Despite the widespread clinical use, CVAD support has been associated with gastrointestinal bleeding, hemorrhagic strokes, and aortic valve insufficiency. Speculation that diminished arterial pressure pulsatility associated with continuous flow devices may be contributing to these complications has sparked much debate over CVAD support. Studies comparing pulsatile flow and continuous flow (CF) support have presented conflicting findings, and the relevance to CVAD as DT is uncertain due to variations in device operation, support duration, and the criteria used to quantify pulsatility. Currently, there is interest in developing control algorithms for CVAD to increase the delivered pulsatility as a strategy to mitigate adverse event risks associated with CVAD therapy. There may also be the added benefit of specific control strategies for managing CVAD therapy, potentially improving the rate of myocardial recovery and successful weaning of mechanical circulatory support.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты