DOI 10.1016/j.athoracsur.2013.04.085
Guidelines for sizing pericardium for aortic valve leaflet grafts.
Аннотация
Surgical repair of the aortic valve with the use of leaflet grafts made from pericardium has been shown to be a viable option, particularly in children, in whom valve replacement has strong disadvantages. We present guidelines for sizing treated autologous pericardium to fabricate a leaflet graft for single-leaflet replacement. Both our clinical experience and experimental evidence indicate that effective repairs are best achieved by use of a semicircular graft with a diameter 10% to 15% greater than the sinotubular junction diameter in diastole. We also provide a simple formula to allow adjustment of these guidelines to account for variations in valve geometry and tissue properties.