DOI 10.1002/ccd.25005

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2013

Three-dimensional multidetector computed tomography versus conventional 2-dimensional transesophageal echocardiography for annular sizing in transcatheter aortic valve replacement: Influence on postprocedural paravalvular aortic regurgitation.

Hansson Nicolaj C, Thuesen Leif, Hjortdal Vibeke E, Leipsic Jonathon, Andersen Henning R, Poulsen Steen H, Webb John G, Christiansen Evald H, Rasmussen Lars E, Krusell Lars R, Terp Kim, Klaaborg Kaj E, Tang Mariann, Lassen Jens F, Bøtker Hans E, Nørgaard Bjarne L

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

In transcatheter aortic valve replacement (TAVR), the influence of aortic annular assessment with either multidetector computed tomography (MDCT) or conventional transesophageal echocardiography (TEE) on the incidence of postprocedural paravalvular aortic regurgitation (PAR) was evaluated.

BACKGROUND

PAR remains a major limitation in TAVR. Appropriate selection of transcatheter heart valve (THV) size is crucial to prevent PAR.

METHODS

Outcomes following TAVR with a balloon-expandable THV were compared in two retrospective cohorts identified according to whether THV size selection was based on TEE (study group 1, n = 80) or MDCT (study group 2, n = 58).

RESULTS

The two study groups were comparable with regard to baseline clinical, risk score, and echocardiographic characteristics. The incidence of moderate/severe PAR was lower in study group 2 than in group 1, 8.6% versus 28.8% (P < 0.01). The difference between the THV nominal diameter and MDCT annular diameter was predictive of moderate/severe PAR (AUC 0.84; 95% CI: 0.72-0.92). Neither age, gender, body mass index, annular eccentricity index, aortic valve calcification nor the difference between the THV diameter and the TEE annular diameter predicted postprocedural PAR. Increased THV oversizing relative to the MDCT mean annular diameter was associated with reduced severity of PAR. No difference in perprocedural complications between two study groups was observed.

CONCLUSION

MDCT-based annular sizing in TAVR significantly reduces postprocedural PAR, and THV oversizing appears pivotal in this aspect. Further delineation of the optimal degree of THV oversizing is needed.

Для цитирования: Hansson Nicolaj C, Thuesen Leif, Hjortdal Vibeke E, Leipsic Jonathon, Andersen Henning R, Poulsen Steen H, Webb John G, Christiansen Evald H, Rasmussen Lars E, Krusell Lars R, Terp Kim, Klaaborg Kaj E, Tang Mariann, Lassen Jens F, Bøtker Hans E, Nørgaard Bjarne L. Three-dimensional multidetector computed tomography versus conventional 2-dimensional transesophageal echocardiography for annular sizing in transcatheter aortic valve replacement: Influence on postprocedural paravalvular aortic regurgitation.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2013. DOI: 10.1002/ccd.25005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.