MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2013.01.138

Медицинская библиотека · JACC. Cardiovascular interventions · 2013

Transapical implantation of a second-generation transcatheter heart valve in patients with noncalcified aortic regurgitation.

Seiffert Moritz, Diemert Patrick, Koschyk Dietmar, Schirmer Johannes, Conradi Lenard, Schnabel Renate, Blankenberg Stefan, Reichenspurner Hermann, Baldus Stephan, Treede Hendrik

СтатьяVideo-Audio MediaПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study sought to report on the feasibility and early results of transcatheter aortic valve implantation employing a second-generation device in a series of patients with pure aortic regurgitation.

BACKGROUND

Efficacy and safety of transcatheter aortic valve implantation in patients with calcific aortic stenosis and high surgical risk has been demonstrated. However, experience with implantation for severe noncalcified aortic regurgitation has been limited due to increased risk for valve dislocation or annular rupture.

METHODS

Five patients (mean age: 66.6 ± 7 years) underwent transapical implantation of a JenaValve (JenaValve Technology GmbH, Munich, Germany) transcatheter heart valve for moderate to severe, noncalcified aortic regurgitation. All patients were considered high risk for surgical aortic valve replacement after evaluation by an interdisciplinary heart team (logistic EuroSCORE [European System for Cardiac Operative Risk Evaluation] range 3.1% to 38.9%). Procedural and acute clinical outcomes were analyzed.

RESULTS

Implantation was successful in all cases without relevant remaining aortic regurgitation or signs of stenosis in any of the patients. No major device- or procedure-related adverse events occurred and all 5 patients were alive with improved exercise tolerance at 3-month follow-up.

CONCLUSIONS

Noncalcified aortic regurgitation continues to be a challenging pathology for transcatheter aortic valve implantation due to the risk for insufficient anchoring of the valve stent within the aortic annulus. This report provides first evidence that the JenaValve prosthesis may be a reasonable option in these specific patients due to its unique stent design, clipping the native aortic valve leaflets, and offering promising early results.

Для цитирования: Seiffert Moritz, Diemert Patrick, Koschyk Dietmar, Schirmer Johannes, Conradi Lenard, Schnabel Renate, Blankenberg Stefan, Reichenspurner Hermann, Baldus Stephan, Treede Hendrik. Transapical implantation of a second-generation transcatheter heart valve in patients with noncalcified aortic regurgitation.. JACC. Cardiovascular interventions. 2013. DOI: 10.1016/j.jcin.2013.01.138.