Radiofrequency Ablation of the Diseased Human Left Ventricle: Biophysical and Electrogram-Based Analysis.

Bates Alexander P, Paisey John, Yue Arthur, Banks Phil, Roberts Paul R, Ullah Waqas

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Predictors of effective ablation lesion delivery in the human left ventricle are not established, particularly in scar. Impedance drop and electrogram (EGM) attenuation are potential surrogates to assess this.

OBJECTIVES

This study sought to establish the relationships between ablation index (AI) and force-time integral (FTI) with impedance drop and EGM attenuation in the human left ventricle.

METHODS

Patients undergoing ventricular tachycardia ablation were recruited. EGMs were collected preablation and postablation, with impedance, AI, and FTI measured during. Based on preablation bipolar voltage, myocardium was adjudged a low-voltage myocardium (LVM) (<0.50 mV), intermediate-voltage myocardium (IVM) (0.51-1.50 mV), and normal-voltage myocardium (NVM) (>1.50 mV). Relationships between these parameters were explored.

RESULTS

A total of 402 ablations were analyzed in 15 patients. The percent impedance drop correlated with AI and FTI (P < 0.0005; repeated-measures correlation coefficient: 0.54 and 0.44, respectively), a relationship that became weaker with increased myocardial fibrosis, (repeated-measures correlation coefficient for NVM, IVM, and LVM, AI: 0.67, 0.60, and 0.52, respectively; FTI: 0.59, 0.51, and 0.42, respectively). The curve between AI/FTI and impedance drop plateaued at 763 AI and 713 gram-seconds, an impedance drop of 7.5%. Shallower curves occurred progressively from NVM to LVM (P < 0.0005). Mixed models demonstrated that AI and FTI had a greater effect on impedance drop than myocardial fibrosis, drift, or orientation, (standardized β: 0.54 and 0.48, respectively). EGMs were attenuated with ablation (29.3%; IQR: 4.4%-53.3%; P < 0.0005), but attenuation did not correlate with AI or FTI.

CONCLUSIONS

On biophysical analysis, ablation beyond an AI of 763 and FTI of 713 gs offers minimal additional efficacy on average. Fibrosis blunts ablation efficacy. AI is a stronger correlate with impedance drop than FTI. EGM attenuation does not correlate with ablation parameters. (Late Potentials and Ablation Index in Ventricular Tachycardia Ablation; NCT03437408).

Для цитирования: Bates Alexander P, Paisey John, Yue Arthur, Banks Phil, Roberts Paul R, Ullah Waqas. Radiofrequency Ablation of the Diseased Human Left Ventricle: Biophysical and Electrogram-Based Analysis.. JACC. Clinical electrophysiology. 2022. DOI: 10.1016/j.jacep.2022.10.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.