Integration of 3D nuclear imaging in 3D mapping system for ventricular tachycardia ablation in patients with implanted devices: Perfusion/voltage retrospective assessment of scar location.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Thibault Bernard, Richer Louis-Philippe, McSpadden Luke C, Ryu Kyungmoo, Aguilar Martin, Cadrin-Tourigny Julia, Tadros Rafik, Mondésert Blandine, Rivard Léna, Dyrda Katia, Dubuc Marc, Macle Laurent, Talajic Mario, Khairy Paul, Guerra Peter G, Roy Denis, Grégoire Jean, Harel François
Аннотация
BACKGROUND: The identification of low-voltage proarrhythmic areas for catheter ablation of scar-mediated ventricular tachycardia (VT) remains challenging. Integration of myocardial perfusion imaging (single-photon emission computed tomography/computed tomography; SPECT/CT) and electroanatomical mapping (EAM) may improve delineation of the arrhythmogenic substrate.
OBJECTIVE: To assess the feasibility of SPECT/CT image integration with voltage maps using the EnSite Precision system (Abbott) in patients undergoing scar-mediated VT ablation.
METHODS: Patients underwent SPECT/CT imaging prior to left ventricular (LV) EAM with the EnSite Precision mapping system. The SPECT/CT, EAM data, and ablation lesions were retrospectively co-registered in the EnSite Precision system and exported for analysis. Segmental tissue viability scores were calculated based on SPECT/CT perfusion and electrogram bipolar voltage amplitude. Concordance, specificity, and sensitivity between the 2 modalities as well as the impact of SPECT/CT spatial resolution were evaluated.
RESULTS: Twenty subjects (95% male, 67 ± 7 years old, left ventricular ejection fraction 36% ± 11%) underwent EAM and SPECT/CT integration. A concordance of 70% was found between EAM and SPECT/CT for identification of cardiac segments as scar vs viable, with EAM showing a 68.5% sensitivity and 76.4% specificity when using SPECT/CT as a gold standard. Projection on low-resolution 3D geometries led to an average decrease of 38% ± 22% of the voltage points used.
CONCLUSION: The study demonstrated the feasibility of integrating SPECT/CT with EAM performed retrospectively for characterization of anatomical substrates during VT ablation procedures.
OBJECTIVE: To assess the feasibility of SPECT/CT image integration with voltage maps using the EnSite Precision system (Abbott) in patients undergoing scar-mediated VT ablation.
METHODS: Patients underwent SPECT/CT imaging prior to left ventricular (LV) EAM with the EnSite Precision mapping system. The SPECT/CT, EAM data, and ablation lesions were retrospectively co-registered in the EnSite Precision system and exported for analysis. Segmental tissue viability scores were calculated based on SPECT/CT perfusion and electrogram bipolar voltage amplitude. Concordance, specificity, and sensitivity between the 2 modalities as well as the impact of SPECT/CT spatial resolution were evaluated.
RESULTS: Twenty subjects (95% male, 67 ± 7 years old, left ventricular ejection fraction 36% ± 11%) underwent EAM and SPECT/CT integration. A concordance of 70% was found between EAM and SPECT/CT for identification of cardiac segments as scar vs viable, with EAM showing a 68.5% sensitivity and 76.4% specificity when using SPECT/CT as a gold standard. Projection on low-resolution 3D geometries led to an average decrease of 38% ± 22% of the voltage points used.
CONCLUSION: The study demonstrated the feasibility of integrating SPECT/CT with EAM performed retrospectively for characterization of anatomical substrates during VT ablation procedures.