Development of an automatic multiplanar reconstruction processing method for head computed tomography.
Аннотация
BACKGROUND
Head computed tomography (CT) is a commonly used imaging modality in radiology facilities. Since multiplanar reconstruction (MPR) processing can produce different results depending on the medical staff in charge, there is a possibility that the antemortem and postmortem images of the same person could be assessed and identified differently.
OBJECTIVE
To propose and test a new automatic MPR method in order to address and overcome this limitation.
METHODS
Head CT images of 108 cases are used. We employ the standardized transformation of statistical parametric mapping 8. The affine transformation parameters are obtained by standardizing the captured CT images. Automatic MPR processing is performed by using this parameter. The sphenoidal sinus of the orbitomeatal cross section of the automatic MPR processing of this study and the conventional manual MPR processing are cropped with a matrix size of 128×128, and the value of zero mean normalized correlation coefficient is calculated.
RESULTS
The computed zero mean normalized cross-correlation coefficient (Rzncc) of≥0.9, 0.8≤Rzncc < 0.9 and 0.7≤Rzncc < 0.8 are achieved in 105 cases (97.2%), 2 cases (1.9%), and 1 case (0.9%), respectively. The average Rzncc was 0.96±0.03.
CONCLUSION
Using the proposed new method in this study, MPR processing with guaranteed accuracy is efficiently achieved.