DOI 10.1016/j.carrev.2019.06.015

Медицинская библиотека · Cardiovascular revascularization medicine : including molecular interventions · 2019

Impact of BASILICA on Sinus and Neo-Sinus Hemodynamics after Valve-in-Valve with and without Coronary Flow.

Hatoum Hoda, Maureira Pablo, Lilly Scott, Dasi Lakshmi Prasad

СтатьяПоддержка исследования: вне правительства СШАVideo-Audio MediaПеревод пока не подготовлен

Аннотация

BACKGROUND/PURPOSE: This study aims at evaluating the impact of BASILICA on neo-sinus and sinus hemodynamics with and without coronary flow. Leaflet thrombosis after valve-in-valve (ViV) may compromise not only leaflet mobility but also affect valve durability and performance.

METHODS/MATERIALS: In a 23 mm transparent surgical aortic valve model, a 23 mm Edwards SAPIEN 3 and a 26 mm Medtronic Evolut were deployed before and after leaflet laceration, in models with and without coronary flow. Neo-sinus and sinus hemodynamics were evaluated in the aortic position of a pulse duplicator and particle image velocimetry was performed in order to quantify sinus flow hemodynamics along with sinus and neo-sinus washout.

RESULTS

BASILICA-type leaflet laceration procedure led to (a) an increase in the velocities in the sinus and the neo-sinus by 50% for Evolut ViV with and without coronary flow, 70% for non-coronary SAPIEN 3 ViV and 10% for coronary SAPIEN 3 ViV, and (b) an improvement in overall washout up to 2 cycles in the neo-sinus and 0.5 cycles in the sinus.

CONCLUSIONS

A BASILICA-type leaflet laceration approach may improve sinus and neo-sinus hemodynamics through decreasing flow stasis and enabling less confined blood flow. BASILICA confers coronary sinus flow patterns to the non-coronary sinus.

Для цитирования: Hatoum Hoda, Maureira Pablo, Lilly Scott, Dasi Lakshmi Prasad. Impact of BASILICA on Sinus and Neo-Sinus Hemodynamics after Valve-in-Valve with and without Coronary Flow.. Cardiovascular revascularization medicine : including molecular interventions. 2019. DOI: 10.1016/j.carrev.2019.06.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.