MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/neurintsurg-2015-011861

Медицинская библиотека · Journal of neurointerventional surgery · 2015

Mechanical thrombectomy with the Trevo ProVue device in ischemic stroke patients: does improved visibility translate into a clinical benefit?

Kabbasch Christoph, Mpotsaris Anastasios, Chang De-Hua, Hiß Sonja, Dorn Franziska, Behme Daniel, Onur Oezguer, Liebig Thomas

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the efficacy and safety of the Trevo ProVue (TPV) stent retriever in stroke patients with large artery occlusions, with particular attention to the full structural radiopacity of the TPV.

MATERIALS AND

METHODS

Case files and images of TPV treatments were reviewed for clinical and technical outcome data, including revascularization rates, device and procedure related complications, and outcome at discharge and after 90 days.

RESULTS

76 patients were treated with TPV. Mean National Institutes of Health Stroke Scale (NIHSS) score was 18 and 68% had additional intravenous thrombolysis. 63 occlusions were in the anterior circulation: 44 M1 (58%), 8 M2 (11%), 8 internal carotid artery-terminus (11%), 2 internal carotid artery- left (3%), 1 A2 (1%), and 13 vertebrobasilar (17%). 58 of 76 (76%) were solely treated with TPV; the remainder were treated with additional stent retrievers. Mean number of passes in TPV only cases was 2.2 (SD 1.2). In rescue cases, 3.2 (SD 2.2) passes were attempted with the TPV followed by 2.6 rescue device passes (SD 2). TPV related adverse events occurred in 4/76 cases (5%) and procedural events in 6/76 cases (8%). Mean procedural duration was 64 min (SD 42). Thrombolysis in Cerebral Infarction (TICI) 2b/3 recanalization was achieved in 69/76 patients (91%), including 50% TICI 3. Of 56 survivors (74%), 37 (49%) showed a favorable outcome at 90 days (Solitaire With the Intention for Thrombectomy trial criteria), statistically associated with age, baseline NIHSS, onset to revascularization time, and TICI 2b-3 reperfusion. TPV radiopacity allowed for visual feedback, changing the methodology of stent retriever use in 44/76 cases (58%).

CONCLUSIONS

Neurothrombectomy with TPV is feasible, effective, and safe. The recanalization rate compares favorably with reported data in the literature. Improved structural radiopacity may facilitate neurothrombectomy or influence the course of action during retrieval.

Для цитирования: Kabbasch Christoph, Mpotsaris Anastasios, Chang De-Hua, Hiß Sonja, Dorn Franziska, Behme Daniel, Onur Oezguer, Liebig Thomas. Mechanical thrombectomy with the Trevo ProVue device in ischemic stroke patients: does improved visibility translate into a clinical benefit?. Journal of neurointerventional surgery. 2015. DOI: 10.1136/neurintsurg-2015-011861.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.