High-grade, non-flow-limiting dissections do not negatively impact long-term outcome after paclitaxel-coated balloon angioplasty: an additional analysis from the THUNDER study.

Tepe Gunnar, Zeller Thomas, Schnorr Beatrix, Claussen Claus D, Beschorner Ulrich, Brechtel Klaus, Scheller Bruno, Speck Ulrich

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the impact of using paclitaxel-coated balloons (PCB) on outcome after post-angioplasty dissection in femoropopliteal arteries.

METHODS

The angiograms obtained in the THUNDER study (ClinicalTrials.gov identifier NCT00156624) were analyzed to compare degrees of dissection and angiographic parameters between the control (uncoated balloons, n=43) and treatment (PCBs, n=43) groups before and after the intervention and at 6-month follow-up. Furthermore, target lesion revascularizations (TLR) were documented up to 2 years.

RESULTS

In each group, 24 (56%) patients had a dissection after the intervention. At the 6-month follow-up, patients with dissection of any grade after treatment with PCBs had significantly less late lumen loss (0.4 mm) than patients with dissection after treatment with uncoated balloons (1.9 mm, p=0.001) and a lower degree of stenosis (20% vs. 51%, respectively; p=0.003). Patients with severe dissection (grades C, D, or E) especially seemed to benefit from the PCBs, with late lumen loss of 0.4 mm vs. 2.4 mm for controls (p=0.05). The binary restenosis rate was also markedly lower in the PCB group (20%) than in the uncoated group (55%, p=0.02). In the 2-year follow-up, TLR was performed in 56% of patients in the control group compared to 10% of patients in the PCB group (p=0.002).

CONCLUSION

The results of this subgroup analysis suggest that patients with dissection following treatment with a paclitaxel-coated balloon have a very acceptable outcome and stent implantation is not necessary as long as the dissection does not result in acute flow limitation.

Для цитирования: Tepe Gunnar, Zeller Thomas, Schnorr Beatrix, Claussen Claus D, Beschorner Ulrich, Brechtel Klaus, Scheller Bruno, Speck Ulrich. High-grade, non-flow-limiting dissections do not negatively impact long-term outcome after paclitaxel-coated balloon angioplasty: an additional analysis from the THUNDER study.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2013. DOI: 10.1583/13-4392r.1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.