m+Medelevator.RU
← Поиск публикаций
The Journal of prosthetic dentistry · 2013

Fracture resistance of soft tissue level implants after cyclic loading and external modification.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

DiPede Louis, Alhashim Abdulmohsin, Vaidyanathan T K, Flinton Robert

Аннотация

STATEMENT OF PROBLEM: The exposure of the implant collar in the esthetic zone is a challenging complication of implant dentistry. The preparation of the exposed implant collar to create room for restorative material may decrease the strength of the abutment/implant assembly.

PURPOSE: The purpose of this in vitro study was to evaluate the effect on the failure load of preparing a 0.5-mm chamfer finish line on an implant collar at various apical depths.

MATERIAL AND METHODS: Thirty solid abutments and 30 implants were screwed into a customized metal mold at a 30-degree angle to the vertical axis. An intact group (IN) (control) (n=10) had no preparation, experimental group P1 had a 0.5-mm chamfer margin placed 1 mm apical to the implant/abutment interface, and experimental group P2 had the margin placed 2 mm apical. The specimens were subjected to a fatigue loading of 200 N for 1 000 000 cycles. Unfractured specimens were then loaded until failure. Mean fracture load (MFL) values were compared with 1-way ANOVA, and the Tukey test was used to identify differences among groups (α=.05).

RESULTS: One specimen from both the IN and P1 groups and 2 specimens from the P2 group failed after fatigue loading. Significant differences in the MFL values were found between the IN group (3825) and P1 group (2944), P<.001 and between the IN group and the P2 group (2553), P<.001.

CONCLUSIONS: The mean load required to fracture the abutment/implant assembly decreased significantly after various amounts of implant collar preparation.