MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/0194599820982635

Медицинская библиотека · Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery · 2021

Laryngeal Force Sensor for Suspension Microlaryngoscopy: A Prospective Controlled Trial.

Feng Allen L, Puka Elefteria, Ciaramella Alex, Rao Vishwanatha M, Wang Tiffany V, Naunheim Matthew R, Song Phillip C

Контролируемое клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The laryngeal force sensor (LFS) provides real-time force data for suspension microlaryngoscopy. This study investigates whether active use of the LFS can prevent the development of complications.

STUDY

DESIGN

Prospective controlled trial.

SETTING

Academic tertiary center.

METHODS

The LFS and custom software were developed to track intraoperative force metrics. A consecutive series of 100 patients had force data collected with operating surgeons blinded to intraoperative readings. The subsequent 100 patients had surgeons actively use the LFS monitoring system. Patients were prospectively enrolled, completing pre- and postoperative surveys to assess the development of tongue pain, paresthesia, paresis, dysgeusia, or dysphagia.

RESULTS

On univariate analysis, the active monitoring group had lower total impulse (P < .001) and fewer extralaryngeal complications (P < .01). On multiple logistic regression, maximum force (odds ratio [OR], 1.08; 95% CI, 1.01-1.16; P = .02) was a significant predictive variable for the development of postoperative complications. Similarly, active LFS monitoring showed a 29.1% (95% CI, 15.7%-42.4%; P < .001) decrease in the likelihood of developing postoperative complications. These effects persisted at the first postoperative visit for maximum force (P = .04) and active LFS monitoring (P = .01). Maximum force (OR, 1.11; 95% CI, 1.04-1.18; P < .01) and active LFS monitoring (16.6%; 95% CI, 2.7%-30.5%; P = .02) were also predictive for the development of an abnormal 10-item Eating Assessment Tool score. These effects also persisted at the first postoperative visit for maximum force (P = .01) and active LFS monitoring (P = .01).

CONCLUSION

Maximum force is predictive of the development postoperative complications. Active monitoring with the LFS is able to mitigate these forces and prevent postoperative complications.

LEVEL OF EVIDENCE: 2.

Для цитирования: Feng Allen L, Puka Elefteria, Ciaramella Alex, Rao Vishwanatha M, Wang Tiffany V, Naunheim Matthew R, Song Phillip C. Laryngeal Force Sensor for Suspension Microlaryngoscopy: A Prospective Controlled Trial.. Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery. 2021. DOI: 10.1177/0194599820982635.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.