Clinica chimica acta; international journal of clinical chemistry · 2017
The standardization of 5 immunoassays for anti-Toxoplasma immunoglobulin G(IgG).
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zhang Kuo, Lin Guigao, Han Yanxi, Li Jinming
Аннотация
BACKGROUND: Quantitative immunoassays to detect IgG antibodies are the most commonly used tests for diagnosing toxoplasmosis. We investigated the current state of standardization of quantitative immunoassays used to measure anti-Toxoplasma IgG levels.
METHODS: Four fully automated immunoassays (Architect i4000ISR, Immulite 2000 Xpi, Siemens; Liaison, DiaSorin; Cobas e601, Roche) and one manual immunoassay (ELISA classic Toxo IgG, Virion Serion) were performed on the following: individual patient serum samples, the WHO international standards, control samples, and calibrators provided by 5 immunoassay manufacturers. Statistical analysis was used to illustrate the results.
RESULTS: No perfect correlation (slope=1.0) was found between any 2 assays. Large differences in anti-Toxoplasma IgG titers were observed among the 5 immunoassays using serum samples from individual patients. Using IS 01/600 as a calibrator minimized the inter-assay variability of anti-Toxoplasma IgG values CONCLUSIONS: There is still significant effort needed towards standardization of anti-Toxoplasma IgG quantitative immunoassays.
METHODS: Four fully automated immunoassays (Architect i4000ISR, Immulite 2000 Xpi, Siemens; Liaison, DiaSorin; Cobas e601, Roche) and one manual immunoassay (ELISA classic Toxo IgG, Virion Serion) were performed on the following: individual patient serum samples, the WHO international standards, control samples, and calibrators provided by 5 immunoassay manufacturers. Statistical analysis was used to illustrate the results.
RESULTS: No perfect correlation (slope=1.0) was found between any 2 assays. Large differences in anti-Toxoplasma IgG titers were observed among the 5 immunoassays using serum samples from individual patients. Using IS 01/600 as a calibrator minimized the inter-assay variability of anti-Toxoplasma IgG values CONCLUSIONS: There is still significant effort needed towards standardization of anti-Toxoplasma IgG quantitative immunoassays.