Automated long axial field of view PET image processing and kinetic modelling with the TurBO toolbox.

Tuisku Jouni, Palonen Santeri, Kärpijoki Henri, Latva-Rasku Aino, Tuomola Nelli, Harju Harri, Nesterov Sergey V, Oikonen Vesa, Iida Hidehiro, Teuho Jarmo, Han Chunlei, Karjalainen Tomi, Kirjavainen Anna K, Rajader Johan, Klén Riku, Nuutila Pirjo, Knuuti Juhani, Nummenmaa Lauri

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Long axial field of view (LAFOV) PET imaging requires extensive automation due to the large number of target tissues. Therefore, we introduce an open-source analysis pipeline (TurBO, Turku total-BOdy) for automated preprocessing and kinetic modelling of LAFOV [15O]H2O and [18F]FDG PET data. TurBO enables efficient, reproducible quantification of tissue perfusion and metabolism at regional- and voxel-levels through automated co-registration, motion correction, CT-based region of interest (ROI) segmentation, image-derived input function (IDIF) extraction, and region-specific kinetic modelling.

METHODS

The pipeline was validated with Biograph Vision Quadra (Siemens Healthineers) LAFOV PET/CT data from 21 subjects scanned with [15O]H2O and 16 subjects scanned with [18F]FDG. Six CT-segmented ROIs (cortical brain gray matter, left iliopsoas muscle, right kidney cortex and medulla, pancreas, spleen and liver) were used to assess different levels of tissue perfusion and glucose metabolism.

RESULTS

Model fits showed high quality with consistent estimates at regional and voxel-levels (R2 > 0.83 for [15O]H2O, R2 > 0.99 for [18F]FDG). Manual and automated IDIFs were in concordance (R2 > 0.74 for [15O]H2O, and R2 > 0.78 for [18F]FDG) with minimal bias (< 4% and < 10%, respectively). Manual and CT-segmented ROIs showed strong agreement (R2 > 0.82 for [15O]H2O and R2 > 0.83 for [18F]FDG). Motion correction had little impact on estimates (R2 > 0.71 for [15O]H2O and R2 > 0.78 for [18F]FDG) compared with uncorrected data.

CONCLUSION

The TurBO pipeline provides fully automated and reliable quantification for LAFOV PET data. It substantially reduces manual workload and enables standardized, reproducible assessment of inter-organ perfusion and metabolism.

Для цитирования: Tuisku Jouni, Palonen Santeri, Kärpijoki Henri, Latva-Rasku Aino, Tuomola Nelli, Harju Harri, Nesterov Sergey V, Oikonen Vesa, Iida Hidehiro, Teuho Jarmo, Han Chunlei, Karjalainen Tomi, Kirjavainen Anna K, Rajader Johan, Klén Riku, Nuutila Pirjo, Knuuti Juhani, Nummenmaa Lauri. Automated long axial field of view PET image processing and kinetic modelling with the TurBO toolbox.. European journal of nuclear medicine and molecular imaging. 2026. DOI: 10.1007/s00259-026-07769-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.