The effect of material assignment in nasal cavity on dose calculation for nasopharyngeal carcinoma (NPC) using Acuros XB.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Cheung Michael L M, Chow Vivian U Y, Kan Monica W K, Chan Anthony T C

Аннотация

PURPOSE: To evaluate the effect of material assignment in nasal cavity on dose calculation for the volumetric modulated arc therapy (VMAT) of nasopharyngeal carcinoma (NPC) using Acuros XB (AXB) algorithm.

METHODS: The VMAT plans of 30 patients with NPC were calculated using AXB with material auto-assignment of nasal cavity to lung and reassignment to air respectively. The doses to the planning target volumes (PTVs) overlapping with nasal cavity with material auto-assignment of lung (AXB_Lung) were compared to the values obtained when nasal cavity was reassigned to air (AXB_Air) under the dose-to-medium (Dm ) reporting mode of AXB.

RESULTS: For dose calculated under AXB_Lung, the D98% , D2% , and Dmean of the PTV69.96 _Air Cavity (PTV of prescription dose 69.96 Gy overlapping with nasal cavity) were on average 16.1%, 1.6%, and 8.6% larger than that calculated under AXB_Air, respectively. Up to 19.5% difference in D98% , 3% difference in D2% , and 11.2% difference in Dmean were observed in the worst cases for PTV69.96 . Similar trend was observed for the PTV5940 _Air Cavity, in which the D98% , D2% , and Dmean calculated under AXB_Lung were on average 14.7%, 2.5%, and 10.2% larger than that calculated under AXB_Air, respectively. In the worst cases, the difference observed in D98% , D2% , and Dmean could be up to 17.7%, 4.5%, and 12.7%, respectively.

CONCLUSIONS: Significant dose difference calculated by AXB between the material assignment of lung and air in nasal cavity for NPC cases might imply the possibility of underdosage to the PTVs that overlap with inhomogeneity. Therefore, attention should be put to ensure that accurate material assignment for dose calculation under AXB such that optimal dosage was given for tumor control.