Use of capillary electrophoresis for accurate determination of CAG repeats causing Huntington disease. An oligonucleotide design avoiding shadow bands.

Blanco Sonia, Suarez Antonio, Gandia-Pla Sandra, Gómez-Llorente Carolina, Antúnez Adelaida, Gómez-Capilla Jose Antonio, Fárez-Vidal M Esther

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Huntington disease (HD) is a neurodegenerative disorder associated with the expansion of a polymorphic trinucleotide CAG repeat in the HD gene. We have developed an assay to accurately determine CAG repeats that combines a novel oligonucleotide design and the resolution of capillary electrophoresis. A mismatch in the second nucleotide from the 3' end enhanced specificity by avoiding mispriming and diminishing shadow bands and artifactual PCR products. The coupling of capillary electrophoresis analysis with the assay added the advantages of accuracy, high resolution, semi-automation, rapid analysis and low sample consumption. Analysis of 200 chromosomes in the Spanish population sample studied (control group) gave a peak frequency for 16 CAG repeats and of 7 triplets for CCG repeats. Diagnosis of HD was confirmed in 22 of 34 individuals with a range of CAG repeats from 39 to 52. Predictive testing was also carried out for 19 relatives of the HD families diagnosed at our laboratory. The method proposed in this article provides an accurate sizing of DNA repeats that can be applied to the analysis of DNA size-related disorders.

Для цитирования: Blanco Sonia, Suarez Antonio, Gandia-Pla Sandra, Gómez-Llorente Carolina, Antúnez Adelaida, Gómez-Capilla Jose Antonio, Fárez-Vidal M Esther. Use of capillary electrophoresis for accurate determination of CAG repeats causing Huntington disease. An oligonucleotide design avoiding shadow bands.. Scandinavian journal of clinical and laboratory investigation. 2008. DOI: 10.1080/00365510801915171.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.