MedElevator Войти в исследования
← Поиск публикаций

DOI 10.21454/rjaic.7518.252.sch

Медицинская библиотека · Romanian journal of anaesthesia and intensive care · 2018

Tracheal intubation with channeled vs. non-channeled videolaryngoscope blades.

Biro Peter, Schlaepfer Martin

СтатьяПеревод пока не подготовлен

Аннотация

STUDY

OBJECTIVE

Videolaryngoscopes can be fitted either with channeled or non-channeled blades, which may result in a different performance and success of tracheal intubation. We investigated the characteristics of the two different blade types of the commercially available KingVision™ videolaryngoscope.

DESIGN

A prospective, randomized, single center investigation study in a urological operation unit of a tertiary hospital.

SUBJECTS AND

METHODS

Forty adult patients undergoing elective urological surgery in general anaesthesia with tracheal intubation were randomly allocated into group 1 (channeled videolaryngoscopy, n = 20) and group 2 (non-channeled videolaryngoscopy, n = 20). We measured the times from laryngoscope insertion to recognize the glottis and to conclude tracheal intubation. The number of laryngoscopy/intubation attempts and the degree of visual glottis exposure on a visual analog scale from 0 (glottis not visible) to 10 (glottis fully visible) was assessed. The lowest SpO2 value during airway management was recorded.

RESULTS

There was no statistically significant difference in biometric data between the 2 groups. The time from the laryngoscope insertion to glottis recognition with the non-channeled blades was 5 (4-8) s as compared to the channeled ones with 11 (7-14) s (median and range; p = 0.01). Intubation duration was shorter with the channeled blades 17 (12-27) s vs. 29 (25-51) s (median and range; p < 0.001). Number of laryngoscopy/intubation attempts, grades for glottis visibility, intubation difficulty were not different. The lowest SpO2 was 98% in both groups.

CONCLUSIONS

Videolaryngoscopic glottis recognition time was longer and the total time to secure the airway was shorter with the channeled blades.

Для цитирования: Biro Peter, Schlaepfer Martin. Tracheal intubation with channeled vs. non-channeled videolaryngoscope blades.. Romanian journal of anaesthesia and intensive care. 2018. DOI: 10.21454/rjaic.7518.252.sch.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.