LMA Protector™ Airway: first experience with a new second generation laryngeal mask.

Eckardt Fabian, Engel Jörg, Mann Sacha T, Müller Matthias, Zajonz Thomas, Koerner Christian M, Sander Michael, Mann Valesco

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The LMA Protector™ Airway (The Laryngeal Mask Company Ltd., Teleflex Incorporated, Athlone, Ireland) is a new supraglottic airway promising a better seal, an improved drainage of gastric secretions and the opportunity of a simplified fiberscopy-guided tracheal intubation. The aim of this study was to present a primary evaluation of the LMA Protector in a clinical setting.

METHODS

After informed consent 50 patients, scheduled for minor/moderate surgery in supine position, were recruited. Pharyngeal seal pressures were examined in neutral position of the patients' head and in maximum passive extension of the neck. Additionally, the fiberscopic view on the glottis was graduated and the feasibility of fiberscope guided tracheal intubation through the device was evaluated.

RESULTS

The median pharyngeal seal pressure of the LMA Protector in neutral position of the head was 34 cmH2O. Passive extension of the neck did not cause a reduction of the pharyngeal seal (median pharyngeal seal pressure: 34.7 cmH2O; P<0.039). The LMA Protector was applicable for fiberscopic tracheal intubation but is not reliable for blind tracheal intubation.

CONCLUSIONS

The LMA Protector provides a high pharyngeal seal. Uncommon for laryngeal masks its pharyngeal seal is not affected by the extension of the patient's neck. As a second generation supraglottic airway which is also suitable for simplified fiberscopic guided tracheal intubation, the LMA Protector could be considered as a supraglottic airway of the third generation.

Для цитирования: Eckardt Fabian, Engel Jörg, Mann Sacha T, Müller Matthias, Zajonz Thomas, Koerner Christian M, Sander Michael, Mann Valesco. LMA Protector™ Airway: first experience with a new second generation laryngeal mask.. Minerva anestesiologica. 2018. DOI: 10.23736/s0375-9393.18.12421-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.