Validation of Hemoglobin and Hematocrit Measurements from a Dialysis Machine Sensor Compared to Laboratory Analysis.

Morisi Niccolò, Ferrarini Marco, Veronesi Laura, Manzini Giovanni, Giovanella Silvia, Alfano Gaetano, Stipo Lucia, Olmeda Fabio, Ligabue Giulia, Virzì Grazia Maria, Di Pinto Valentina, Rovati Luigi, Donati Gabriele

СтатьяПеревод пока не подготовлен

Аннотация

Background

Continuous monitoring of hemoglobin (HB) and hematocrit (HCT) during hemodialysis could improve fluid management and patient safety. The Fresenius 5008 dialysis machine includes an ultrasound-based sensor that estimates HB and HCT values, though its accuracy compared to standard laboratory measurements remains unclear.

Methods

This exploratory observational study assessed the agreement between sensor-derived and laboratory-derived HB and HCT values in 20 patients at the start of hemodiafiltration. A total of 177 paired blood samples were collected.

Results

Sensor values significantly underestimated laboratory HB (9.61 vs. 11.31 g/dL) and HCT (27% vs. 34%) (p < 8 × 10-25). Correlations were strong for both parameters (HB: r = 0.788; HCT: r = 0.876). Regression analyses revealed consistent proportional bias. Applying a fixed correction of +1.69 g/dL for HB and +7.55% for HCT eliminated the statistical differences and reduced intercepts in regression models. Bland-Altman plots confirmed improved agreement post-correction. Albumin levels correlated modestly with error magnitude.

Conclusions

HB and HCT values from the Fresenius 5008 sensor are strongly correlated with laboratory data but are systematically underestimated at treatment start, likely due to hemodilution. Applying fixed correction factors improves accuracy and supports the sensor's use for real-time monitoring.

Для цитирования: Morisi Niccolò, Ferrarini Marco, Veronesi Laura, Manzini Giovanni, Giovanella Silvia, Alfano Gaetano, Stipo Lucia, Olmeda Fabio, Ligabue Giulia, Virzì Grazia Maria, Di Pinto Valentina, Rovati Luigi, Donati Gabriele. Validation of Hemoglobin and Hematocrit Measurements from a Dialysis Machine Sensor Compared to Laboratory Analysis.. Journal of clinical medicine. 2025. DOI: 10.3390/jcm14155242.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.