Clinical application of direct bronchial ultrasound to visualize and determine endobronchial tumor margins for surgical resection.

Клинический случайСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Sarraf Khaled M, Belcher Elizabeth, Price Susanna, Lim Eric

Аннотация

PURPOSE: We describe the first experience of direct bronchial (epi-bronchial) ultrasound to visualize and determine the endobronchial tumor margins for surgical resection.

DESCRIPTION: An ultrasound probe was applied onto the membranous portion of the right main bronchus directly over a pedunculated tumor. The tumor was visualized due to the water content, with a total loss of signal (air-tumor interface) at the tumor edge. A sterile marker was used to outline the air-tumor interface.

EVALUATION: Traditionally, surgical technique involves palpation of the tumor with an incision to inspect the endobronchial lumen determining the position of the pedicle, thus estimating the amount of airway to resect. Using direct bronchial ultrasound, the right main palpated margin was 0.5-cm proximal to the ultrasound margin, which correctly identified the tumor margin. The upper lobe palpated margin was 1 cm proximal to the ultrasound margin, which correctly identified the tumor margin. In the intermediate bronchus, the palpated and ultrasound margin were the same and correct.

CONCLUSIONS: By using the air-tumor interface, epi-bronchial ultrasound scanning can accurately demarcate the base of endobronchial tumors for surgical resection and reconstruction.