m+Medelevator.RU
← Поиск публикаций
Archives of physical medicine and rehabilitation · 2017

Long-Term Performance and User Satisfaction With Implanted Neuroprostheses for Upright Mobility After Paraplegia: 2- to 14-Year Follow-Up.

СтатьяПоддержка NIH: внешняя программаПоддержка правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Triolo Ronald J, Bailey Stephanie Nogan, Foglyano Kevin M, Kobetic Rudi, Lombardo Lisa M, Miller Michael E, Pinault Gilles

Аннотация

OBJECTIVE: To quantify the long-term (>2y) effects of lower extremity (LE) neuroprostheses (NPs) for standing, transfers, stepping, and seated stability after spinal cord injury.

DESIGN: Single-subject design case series with participants acting as their own concurrent controls, including retrospective data review.

SETTING: Hospital-based clinical biomechanics laboratory with experienced (>20y in the field) research biomedical engineers, a physical therapist, and medical monitoring review.

PARTICIPANTS: Long-term (6.2±2.7y) at-home users (N=22; 19 men, 3 women) of implanted NPs for trunk and LE function with chronic (14.4±7.1y) spinal cord injury resulting in full or partial paralysis.

INTERVENTIONS: Technical and clinical performance measurements, along with user satisfaction surveys.

MAIN OUTCOME MEASURES: Knee extension moment, maximum standing time, body weight supported by lower extremities, 3 functional standing tasks, 2 satisfaction surveys, NP usage, and stability of implanted components.

RESULTS: Stimulated knee extension strength and functional capabilities were maintained, with 94% of implant recipients reporting being very or moderately satisfied with their system. More than half (60%) of the participants were still using their implanted NPs for exercise and function for >10min/d on nearly half or more of the days monitored; however, maximum standing times and percentage body weight through LEs decreased slightly over the follow-up interval. Stimulus thresholds were uniformly stable. Six-year survival rates for the first-generation implanted pulse generator (IPG) and epimysial electrodes were close to 90%, whereas those for the second-generation IPG along with the intramuscular and nerve cuff electrodes were >98%.

CONCLUSIONS: Objective and subjective measures of the technical and clinical performances of implanted LE NPs generally remained consistent for 22 participants after an average of 6 years of unsupervised use at home. These findings suggest that implanted LE NPs can provide lasting benefits that recipients value.