MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0000000000003681

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2022

Initial Experience With Two Active Transcutaneous Bone-Anchored Hearing Implants.

Schwam Zachary G, Perez Enrique R, Oh Samuel, Wong Kevin, Fan Caleb, Cosetti Maura K, Wanna George B

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To analyze our institutional experience with two active transcutaneous bone-anchored hearing aids.

STUDY

DESIGN

Retrospective cohort study.

SETTING

Tertiary academic otology-neurotology practice.

PATIENTS: Patients with conductive or mixed hearing losses meeting criteria to receive active transcutaneous bone-anchored hearing aids.

INTERVENTIONS: Implantation with one of two active transcutaneous bone-anchored hearing aids.

OUTCOME MEASURES: Operative time, dural exposure and decompression, use of lifts, implant position, ease of use, qualitative patient satisfaction, complication rates.

RESULTS

Ten patients received Implant 1 and 11 patients received Implant 2. The most common underlying etiologies of hearing loss were chronic suppurative otitis media in 33.3%, atresia/microtia in 23.8%, and cholesteatoma in 23.8%. Average operative times were 99.3 minutes for Implant 1 and 80.9 minutes for Implant 2 ( p = 0.263). Implant 1 required lifts in 60%, dural exposure in 50%, and dural compression in 30%. Overall, placement was difficult in 47.6% of cases. There were no intraoperative complications. There were higher rates of issues with sound quality (27.3% versus 0.0%, p = 0.123) and poor cosmesis (36.4% versus 10.0%, p = 0.360) with Implant 2. Functional gain was not recorded, but all patients derived qualitative benefit from their implant. Three patients had local wound complications that self-resolved or were treated with antibiotics. One patient implanted at an outside institution required explant because of multiple infections.

CONCLUSIONS

There were no statistically significant differences in outcomes comparing Implants 1 and 2; however, Implant 2 had much higher rates of issues with audio quality and poor cosmesis. Placing Implant 1 often required special techniques.

Для цитирования: Schwam Zachary G, Perez Enrique R, Oh Samuel, Wong Kevin, Fan Caleb, Cosetti Maura K, Wanna George B. Initial Experience With Two Active Transcutaneous Bone-Anchored Hearing Implants.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2022. DOI: 10.1097/mao.0000000000003681.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.