DOI 10.1515/cclm-2016-0275

Медицинская библиотека · Clinical chemistry and laboratory medicine · 2017

Evaluation and comparison of three assays for molecular detection of spinal muscular atrophy.

Li Liang, Zhou Wan-Jun, Fang Ping, Zhong Ze-Yan, Xie Jian-Sheng, Yan Ti-Zhen, Zeng Jian, Tan Xu-Hui, Xu Xiang-Min

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Spinal muscular atrophy (SMA) is mainly caused by deletions in SMA-related genes. The objective of this study was to develop gene-dosage assays for diagnosing SMA.

METHODS

A multiplex, quantitative PCR assay and a CNVplex assay were developed for determining the copy number of SMN1, SMN2, and NAIP. Reproducibility and specificity of the two assays were compared to a multiple ligation-dependent probe amplification (MLPA) assay. To evaluate reproducibility, 30 samples were analyzed three times using the three assays. A total of 317 samples were used to assess the specificity of the two assays.

RESULTS

The multiplex quantitative PCR (qPCR) assay had higher reproducibility. Intra-assay CVs were 3.01%-8.52% and inter-assay CVs were 4.12%-6.24%. The CNVplex assay had ratios that were closer to expected (0.49-0.5 for one copy, 1.03-1.0 for two copies, and 1.50-1.50 for three copies). Diagnostic accuracy rates for the two assays were 100%.

CONCLUSIONS

The multiplex qPCR assay was a simple, rapid, and cost-effective method for routine SMA diagnosis and carrier screening. The CNVplex assay could be used to detect SMAs with complicated gene structures. The assays were reliable and could be used as alternative methods for clinical diagnosis of SMA.

Для цитирования: Li Liang, Zhou Wan-Jun, Fang Ping, Zhong Ze-Yan, Xie Jian-Sheng, Yan Ti-Zhen, Zeng Jian, Tan Xu-Hui, Xu Xiang-Min. Evaluation and comparison of three assays for molecular detection of spinal muscular atrophy.. Clinical chemistry and laboratory medicine. 2017. DOI: 10.1515/cclm-2016-0275.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.