DOI 10.1097/cce.0000000000000118

Медицинская библиотека · Critical care explorations · 2020

Personalized Ventilation to Multiple Patients Using a Single Ventilator: Description and Proof of Concept.

Han Jay S, Mashari Azad, Singh Devin, Dianti Jose, Goligher Ewan, Long Michael, Ng William, Wasowicz Marcin, Preiss David, Vesely Alex, Kacmarek Robert, Keshavjee Shaf, Brochard Laurent, Fisher Joseph A, Slutsky Arthur S

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To design and test a ventilator circuit that can be used for ventilation of two or more patients with a single ventilator, while allowing individualization of tidal volume, fractional concentration of oxygen, and positive end-expiratory pressure to each patient, irrespective of the other patient's respiratory system mechanics.

DESIGN

Description and proof of concept studies.

SETTINGS: Respiratory therapy laboratory.

SUBJECTS: Ventilation of mechanical test lungs.

INTERVENTIONS: Following a previously advocated design, we used components readily available in our hospital to assemble two "bag-in-a-box" breathing circuits. Each patient circuit consisted of a flexible bag in a rigid container connected via one-way valve to a test lung, along with an inline positive end-expiratory pressure valve, connected to the ventilator's expiratory limb. Compressed gas fills the bags during "patient" exhalation. During inspiration, gas from the ventilator, in pressure control mode, enters the containers and displaces gas from the bags to the test lungs. We varied tidal volume, "respiratory system" compliance, and positive end-expiratory pressure in one lung and observed the effect on the tidal volume of the other.

MEASUREMENTS AND MAIN

RESULTS

We were able to obtain different tidal volume, dynamic driving pressure, and positive end-expiratory pressure in the two lungs under widely different compliances in both lungs. Complete obstruction, or disconnection at the circuit connection to one test lung, had minimal effect (< 5% on average) on the ventilation to the co-ventilated lung.

CONCLUSIONS

A secondary circuit "bag-in-the-box" system enables individualized ventilation of two lungs overcoming many of the concerns of ventilating more than one patient with a single ventilator.

Для цитирования: Han Jay S, Mashari Azad, Singh Devin, Dianti Jose, Goligher Ewan, Long Michael, Ng William, Wasowicz Marcin, Preiss David, Vesely Alex, Kacmarek Robert, Keshavjee Shaf, Brochard Laurent, Fisher Joseph A, Slutsky Arthur S. Personalized Ventilation to Multiple Patients Using a Single Ventilator: Description and Proof of Concept.. Critical care explorations. 2020. DOI: 10.1097/cce.0000000000000118.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.