DOI 10.1016/j.shj.2026.101084

Медицинская библиотека · Structural heart : the journal of the Heart Team · 2026

Durability Analysis of a Self-Expanding Transcatheter Aortic Valve Through 30 Years of Accelerated Wear Testing.

Attizzani Guilherme, Frazzetto Marco, De Backer Ole, Dvir Danny, Garlepp Vanessa, Kelley Tim, Grubb Kendra J, Tang Gilbert H L

СтатьяПеревод в очереди

Аннотация

•First reported 30-year in vitro accelerated wear testing data for a THV.•All Evolut PRO + valves met ISO hydrodynamic criteria through 20 years of testing.•At least one valve per size maintained performance through 30 years of testing.•Failures manifested in a consistent pattern, suggesting a shared mechanism.•Testing is ongoing until all valves fail.

Для цитирования: Attizzani Guilherme, Frazzetto Marco, De Backer Ole, Dvir Danny, Garlepp Vanessa, Kelley Tim, Grubb Kendra J, Tang Gilbert H L. Durability Analysis of a Self-Expanding Transcatheter Aortic Valve Through 30 Years of Accelerated Wear Testing.. Structural heart : the journal of the Heart Team. 2026. DOI: 10.1016/j.shj.2026.101084.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.