MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/jbm.b.35010

Медицинская библиотека · Journal of biomedical materials research. Part B, Applied biomaterials · 2022

Surface wear in a custom manufactured temporomandibular joint prosthesis.

De Meurechy Nikolas, Aktan Merve Kübra, Boeckmans Bart, Huys Stijn, Verwilghen Denis R, Braem Annabel, Mommaerts Maurice Y

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

The wear of a novel temporomandibular joint (TMJ) prosthesis was evaluated in an animal model. The prosthesis consisted of an additively manufactured titanium alloy (Ti6 Al4 V) mandibular condyle and glenoid fossa created through selective laser melting, with a machined vitamin E-enriched ultra-high molecular weight polyethylene (UHMWPE) surface attached to the fossa. Thirteen TMJ prosthesis were implanted in sheep, six of which had condylar heads coated with HadSat® diamond-like carbon (H-DLC). Euthanasia took place after 288 days, equaling 22 years of human mastication. Linear and volumetric wear analysis of the fossa was performed by optical scanning. The condylar head surfaces were assessed by scanning electron and confocal laser microscopy. The average linear UHMWPE wear, when combined with the coated condyle, was 0.67 ± 0.28 mm (range: 0.34-1.15 mm), not significantly differing (p = .3765, t-test) from the non-coated combination average (0.88 ± 0.41 mm; range: 0.28-1.48 mm). The respective mean volumetric wear volumes were 25.29 ± 11.43 mm3 and 45.85 ± 22.01 mm3 , not significantly differing (p = .1448, t-test). Analysis of the coated condylar surface produced a mean Ra of 0.12 ± 0.04 μm and Sa of 0.69 ± 0.07 μm. The non-coated condylar surface measured a mean Ra of 0.28 ± 0.17 μm and Sa of 2.40 ± 2.08 μm. Both Sa (p = .0083, Mann-Whitney U test) and Ra (p = .0182, Mann-Whitney U test), differed significantly. The prosthesis exhibits acceptable wear resistance and addition of the H-DLC-coating significantly improved long-term condylar surface smoothness.

Для цитирования: De Meurechy Nikolas, Aktan Merve Kübra, Boeckmans Bart, Huys Stijn, Verwilghen Denis R, Braem Annabel, Mommaerts Maurice Y. Surface wear in a custom manufactured temporomandibular joint prosthesis.. Journal of biomedical materials research. Part B, Applied biomaterials. 2022. DOI: 10.1002/jbm.b.35010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.