DOI 10.1109/jtehm.2021.3099971
3D-Printed Ophthalmic-Retrobulbar-Anesthesia Simulator: Mimicking Anatomical Structures and Providing Tactile Sensations.
Аннотация
Objective
A simulator for retrobulbar anesthesia administration mimicking the orbital anatomy and providing tactile sensation is proposed.
Methods
The production process involves 3D modeling of anatomical structures on the basis of computerized tomography (CT) images, printing the models using a 3D printer, and casting the silicone. Twenty ophthalmologists administered retrobulbar anesthesia using the simulator with four different ocular axial lengths (including extreme myopes); the position of the needle tip was evaluated. The effectiveness of this simulator for training was also surveyed.
Results
The proportions of the final location of the needle tip were 59.25%, 36.25%, and 4.5% for the retrobulbar space, peribulbar space, and intraocular space, respectively. Experienced ophthalmologists showed lower complication rates than residents (0.5% vs 8.5%, [Formula: see text]) and agreed that this simulator will help young ophthalmologists advance their anesthesia-administering skills. Discussion/
Conclusion
The 3D-printered simulator for retrobulbar anesthesia was produced and performance was verified. The technology could be used to simulate critical orbital anatomic features and could be used as a training tool for resident ophthalmologists.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.