MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/diagnostics10040191

Медицинская библиотека · Diagnostics (Basel, Switzerland) · 2020

Minimal Cardiac Perforation by Lead Pacemaker Complicated with Pericardial Effusion and Impending Tamponade: Optimal Management with No Pericardiocentesis Driven by Echocardiography.

Caiati Carlo, Pollice Paolo, Truncellito Luigi, Lepera Mario Erminio, Favale Stefano

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

We report the case of a 51-year-old patient who underwent the implantation of a bi-ventricular implantable cardioverter defibrillator (ICD) complicated by a sub-acute right ventricular minimal perforation with pericardial effusion and echocardiographic signs of tamponade. A new echocardiographic plane orientation allowed us to diagnose this condition in emergency and to make the right decision without delay, which consisting in unscrewing the active fixation screw under fluoroscopy guidance, while the pericardiocentesis was postponed. Thanks to the intervention focused on eliminating the cause of the postcardiac injury syndrome, the patient recovered rapidly and ultimately avoided the pericardiocentesis procedure.

Для цитирования: Caiati Carlo, Pollice Paolo, Truncellito Luigi, Lepera Mario Erminio, Favale Stefano. Minimal Cardiac Perforation by Lead Pacemaker Complicated with Pericardial Effusion and Impending Tamponade: Optimal Management with No Pericardiocentesis Driven by Echocardiography.. Diagnostics (Basel, Switzerland). 2020. DOI: 10.3390/diagnostics10040191.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.