Gastrointestinal endoscopy · 2023
A novel blood-sensing capsule for rapid detection of upper GI bleeding: a prospective clinical trial.
СтатьяКлиническое исследованиеСравнительное исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Akiki Karl, Mahmoud Tala, Alqaisieh Mohammad H, Sayegh Lea N, Lescalleet Kristin E, Abu Dayyeh Barham K, Wong Kee Song Louis M, Larson Mark V, Bruining David H, Coelho-Prabhu Nayantara, Buttar Navtej S, Sedlack Robert E, Chandrasekhara Vinay, Leggett Cadman L, Law Ryan J, Rajan Elizabeth, Gleeson Ferga C, Alexander Jeffrey A, Storm Andrew C
Аннотация
BACKGROUND AND AIMS: Upper GI bleeding (UGIB) is a common medical emergency associated with high resource utilization, morbidity, and mortality. Timely EGD can be challenging from personnel, resource, and access perspectives. PillSense (EnteraSense Ltd, Galway, Ireland) is a novel swallowed bleeding sensor for the detection of UGIB, anticipated to aid in patient triage and guide clinical decision-making for individuals with suspected UGIB.
METHODS: This prospective, open-label, single-arm comparative clinical trial of a novel bleeding sensor for patients with suspected UGIB was performed at a tertiary care center. The PillSense system consists of an optical sensor and an external receiver that processes and displays data from the capsule as "Blood Detected" or "No Blood Detected." Patients underwent EGD within 4 hours of capsule administration; participants were followed up for 21 days to confirm capsule passage.
RESULTS: A total of 126 patients were accrued to the study (59.5% male; mean age, 62.4 ± 14.3 years). Sensitivity and specificity for detecting the presence of blood were 92.9% (P = .02) and 90.6% (P < .001), respectively. The capsule's positive and negative predictive values were 74.3% and 97.8%, and positive and negative likelihood ratios were 9.9 and .08. No adverse events or deaths occurred related to the PillSense system, and all capsules were excreted from patients on follow-up.
CONCLUSIONS: The PillSense system is safe and effective for detecting the presence of blood in patients evaluated for UGIB before upper GI endoscopy. It is a rapidly deployed tool, with easy-to-interpret results that will affect the diagnosis and triage of patients with suspected UGIB. (Clinical trial registration number: NCT05385224.).
METHODS: This prospective, open-label, single-arm comparative clinical trial of a novel bleeding sensor for patients with suspected UGIB was performed at a tertiary care center. The PillSense system consists of an optical sensor and an external receiver that processes and displays data from the capsule as "Blood Detected" or "No Blood Detected." Patients underwent EGD within 4 hours of capsule administration; participants were followed up for 21 days to confirm capsule passage.
RESULTS: A total of 126 patients were accrued to the study (59.5% male; mean age, 62.4 ± 14.3 years). Sensitivity and specificity for detecting the presence of blood were 92.9% (P = .02) and 90.6% (P < .001), respectively. The capsule's positive and negative predictive values were 74.3% and 97.8%, and positive and negative likelihood ratios were 9.9 and .08. No adverse events or deaths occurred related to the PillSense system, and all capsules were excreted from patients on follow-up.
CONCLUSIONS: The PillSense system is safe and effective for detecting the presence of blood in patients evaluated for UGIB before upper GI endoscopy. It is a rapidly deployed tool, with easy-to-interpret results that will affect the diagnosis and triage of patients with suspected UGIB. (Clinical trial registration number: NCT05385224.).