Electrical Dyssynchrony on Noninvasive Electrocardiographic Mapping correlates with SAI QRST on surface ECG.
Аннотация
AIM: The goal of this study was to compare associations between clinical and ECG predictors of cardiac resynchronization therapy (CRT) response with electrical dyssynchrony.
METHODS
Body-surface potentials were recorded using a 120-lead system in 4 patients (age 62 ± 12 y, left ventricular ejection fraction (LVEF) 29 ± 5 %; QRS duration 154 ± 19 ms) with post-myocardial infarction scar and left bundle branch block before CRT implantation. A patient-specific heart-torso model derived from MRI with 291 heart-surface nodes was developed. Electrical dyssynchrony index (EDI) was computed as the standard deviation of activation times on the epicardium while uncoupling index (UI) was measured as the difference between the activation times.
RESULTS
QRS duration correlated with mean activation time (r = 0.977; P = 0.023), but did not correlate with EDI or UI. LVEF inversely correlated with activation time at the lowest 20th percentile (r = -0.960; P = 0.040). Sum absolute QRST integral (SAI QRST), measured on orthogonal XYZ ECG, correlated with EDI (r = 0.955; P = 0.045), and characterized late-activated area of the left ventricle.
CONCLUSION
SAI QRST is a measure of electrical dyssynchrony on ECG.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.