DOI 10.1016/j.jcin.2015.09.036

Медицинская библиотека · JACC. Cardiovascular interventions · 2016

Predictors of Permanent Pacemaker Implantations and New-Onset Conduction Abnormalities With the SAPIEN 3 Balloon-Expandable Transcatheter Heart Valve.

Husser Oliver, Pellegrini Costanza, Kessler Thorsten, Burgdorf Christof, Thaller Hannah, Mayr N Patrick, Kasel Albert M, Kastrati Adnan, Schunkert Heribert, Hengstenberg Christian

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study investigates the influence of implantation depth and prosthesis oversizing on conduction abnormalities (CA) and permanent pacemaker implantation (PPI) after SAPIEN 3 (Edwards Lifesciences, Irvine, California) implantation.

BACKGROUND

CA and PPIs are frequent complications after transcatheter aortic valve replacement with a next-generation balloon-expandable transcatheter heart valve (SAPIEN 3). The potential underlying mechanisms are incompletely understood.

METHODS

Of 244 patients treated with SAPIEN 3,208 without a previous pacemaker and 184 without baseline CA were analyzed. We assessed the association of angiographic implantation depth (% of frame height below the annulus) and degree of oversizing with PPI and CA.

RESULTS

New PPI and new or worsened CA or PPI occurred in 16% (34 of 208) and 31% (57 of 184), respectively. Patients requiring PPI had a higher prevalence of atrial fibrillation (44% vs. 24%; p = 0.017), complete right bundle branch block (27% vs. 5%; p = 0.001), and bradycardia (<60 beats/min, 38% vs. 21%; p = 0.034). In patients with new CA or PPI, implantation depth was lower (at septal side: 29 ± 8% vs. 25 ± 7%; p = 0.003), and rate of oversizing was higher (19% [11 of 57] vs. 6% [8 of 126]; p = 0.007). Independent predictors of new or worsened CA or PPI were implantation depth at septal side (odds ratio [OR]: 1.063 [95% confidence interval (CI): 1.017 to 1.110]; p = 0.006 per % of frame below the aortic annulus), oversizing (OR: 3.489 [95% CI: 1.236 to 9.848]; p = 0.018), and QRS duration (OR: 1.033 [95% CI: 1.011 to 1.056]; p = 0.003 per ms).

CONCLUSIONS

Implantation depth and prosthesis oversizing were associated with a higher rate of new CA or PPI using the SAPIEN 3. Thus, avoidance of deep implantation and extreme oversizing may reduce these complications.

Для цитирования: Husser Oliver, Pellegrini Costanza, Kessler Thorsten, Burgdorf Christof, Thaller Hannah, Mayr N Patrick, Kasel Albert M, Kastrati Adnan, Schunkert Heribert, Hengstenberg Christian. Predictors of Permanent Pacemaker Implantations and New-Onset Conduction Abnormalities With the SAPIEN 3 Balloon-Expandable Transcatheter Heart Valve.. JACC. Cardiovascular interventions. 2016. DOI: 10.1016/j.jcin.2015.09.036.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.