MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1161/circep.115.003004

Медицинская библиотека · Circulation. Arrhythmia and electrophysiology · 2015

Left Ventricular Lead Electrical Delay Is a Predictor of Mortality in Patients With Cardiac Resynchronization Therapy.

Roubicek Tomas, Wichterle Dan, Kucera Pavel, Nedbal Pavel, Kupec Jindrich, Sedlakova Jana, Cerny Jan, Stros Jan, Kautzner Josef, Polasek Rostislav

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Electric left ventricular lead position, assessed by the electric delay from the beginning of the QRS complex to the local LV electrogram (QLV), was found in previous studies to be a strong predictor of short-term response to cardiac resynchronization therapy. We hypothesized that suboptimum electric position of the left ventricular lead is associated with an excess of heart failure events and mortality.

METHODS AND

RESULTS

We analyzed the clinical outcome of patients with left bundle branch block or intraventricular conduction delay treated with cardiac resynchronization therapy at our institution during 9 years. Baseline clinical characteristics, QLV/QRS duration (QLV ratio) at cardiac resynchronization therapy implant, and data about heart failure hospitalization and mode of death were collected in 329 patients who were followed for a period of 3.3±1.9 years. Of them, 83 were hospitalized for heart failure and 83 died. Event rates for all-cause mortality, cardiac mortality, noncardiac mortality, heart failure mortality, and sudden death were 25.2%, 14.9%, 10.3%, 12.2%, and 2.1%, respectively. Patients with a QLV ratio ≤0.70 had significantly worse event-free survival for all study end points--hazard ratio, 1.6; 95% confidence interval, 1.0 to 2.4; P=0.05 for heart failure hospitalization; hazard ratio, 2.9; 95% confidence interval, 1.6 to 5.5; P=0.001 for heart failure mortality; hazard ratio, 1.8; 95% confidence interval, 1.1 to 2.7; P=0.01 for cardiac mortality; and hazard ratio, 2.1; 95% confidence interval, 1.2 to 3.7; P=0.01 for all-cause mortality. In multivariable analysis, QLV ratio ≤0.70 remained associated with all study end points.

CONCLUSIONS

Electric left ventricular lead position in cardiac resynchronization therapy patients was a significant predictor of heart failure hospitalization and mortality.

Для цитирования: Roubicek Tomas, Wichterle Dan, Kucera Pavel, Nedbal Pavel, Kupec Jindrich, Sedlakova Jana, Cerny Jan, Stros Jan, Kautzner Josef, Polasek Rostislav. Left Ventricular Lead Electrical Delay Is a Predictor of Mortality in Patients With Cardiac Resynchronization Therapy.. Circulation. Arrhythmia and electrophysiology. 2015. DOI: 10.1161/circep.115.003004.