A novel approach to hearing amplification: audiometric outcomes from Nuance Audio over-the-counter hearing aid glasses.
Аннотация
OBJECTIVE
To evaluate the first audiometric outcomes of Nuance Audio hearing glasses, an over-the-counter solution integrating air conduction amplification technology into standard eyeglass frames, in individuals with mild to moderate hearing loss.
METHODS
Thirty-two adults (mean age 74 years) with symmetric, age-related mild to moderate sensorineural hearing loss were tested. Pure-tone and speech audiometry were performed under unaided and aided free-field conditions. Device settings were adjusted via smartphone app based on audiometric profiles and user preference. Paired t-tests assessed the differences.
RESULTS
Nuance Audio glasses significantly improved hearing thresholds, especially at 4000-6000 Hz, with mean gains of 10-11 dB HL. Speech reception thresholds improved by 7 dB at 50% intelligibility and 6 dB at 100%. Amplification profile preferences matched audiometric slopes and no differences in benefit were found between mild and moderate hearing loss groups.
CONCLUSIONS
Nuance Audio glasses provide clinically relevant improvements in pure-tone and speech thresholds for adults with mild to moderate presbycusis. Their discreet, open-ear design and self-fitting approach address barriers to traditional hearing aids and may support earlier adoption.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.