DOI 10.1007/s00268-023-07107-0

Медицинская библиотека · World journal of surgery · 2023

"In Vitro" Evaluation of Energy-Based Sealing of Graft Side Branches in Bypass Surgery.

Miralles Manuel, Falcón Moisés, Requejo Lucía, Plana Emma, Medina Pilar, Sánchez-Nevárez Ignacio, Clará Albert

СтатьяСравнительное исследованиеОценочное исследованиеПеревод пока не подготовлен

Аннотация

INTRODUCTION

Our objective was to compare the in vitro efficacy of electrothermal bipolar [EB] vessel sealing and ultrasonic harmonic scalpel [HS] versus mechanical interruption, with conventional ties or surgical clips (SC), in sealing saphenous vein (SV) collaterals, during its eventual preparation for bypass surgery.

METHODS

Experimental in vitro study on 30 segments of SV. Each fragment included two collaterals at least 2 mm in diameter. One of them was sealed by ligation with 3/0 silk ties (control) and the other one with EB (n = 10), HS (n = 10) or medium-6 mm SC (n = 10). After incorporation in a closed circuit with pulsatile flow, the pressure was progressively increased until causing rupture. Collateral diameter, burst pressure, leak point, and histological study were recorded.

RESULTS

Burst pressure was higher for SC (1320.20 ± 373.847 mmHg) as compared with EB (942.2 ± 344.9 mmHg, p = 0.065), and especially with HS (637.00 ± 320.61 mmHg, p = 0.0001). No statistically significant difference between EB and HS was found, and bursting always happened at supraphysiological pressures. The leak point for HS was always detected in the sealing zone (10/10), while for EB and SC, it occurred in the sealing zone only in 6/10(60%) and 4/10(40%), respectively (p = 0.015).

CONCLUSIONS

Energy delivery devices showed similar efficacy and safety in sealing of SV side branches. Although bursting pressure was lower than with tie ligature or SC, non-inferiority efficacy was shown at the range of physiological pressures in both, EB and HS. Due to their speed and easy handling, they may be useful in the preparation of the venous graft during revascularization surgery. However, remaining questions about healing process, potential spread of tissue damage and sealing durability, will require further analysis.

Для цитирования: Miralles Manuel, Falcón Moisés, Requejo Lucía, Plana Emma, Medina Pilar, Sánchez-Nevárez Ignacio, Clará Albert. "In Vitro" Evaluation of Energy-Based Sealing of Graft Side Branches in Bypass Surgery.. World journal of surgery. 2023. DOI: 10.1007/s00268-023-07107-0.