American heart journal · 2010
Intraoperative defibrillation threshold testing during implantable cardioverter-defibrillator insertion: do we really need it?
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Calvi Valeria, Dugo Daniela, Capodanno Davide, Arancio Rocco, Di Grazia Angelo, Liotta Claudio, Puzzangara Euglena, Ragusa Antonio, Arestia Alberto, Tamburino Corrado
Аннотация
BACKGROUND: The assessment of defibrillation efficacy using a safety margin of 10 J has long been the standard of care for insertion of implantable cardioverter-defibrillator (ICD), but physicians are concerned about complications related to induction test. Therefore, the need for testing has been recently questioned. The aim of our study was to assess the impact of defibrillation threshold (DFT) testing of ICD on the efficacy of ICD therapy.
METHODS: We analyzed data obtained from follow-up visits of 122 consecutive patients who underwent ICD implantation at our institute from April 1996 to June 2008, with (n = 42) or without (n = 80) DFT testing. Patients in the DFT group were less likely to be men (83.3% vs 96.3%, P < .031) than those in the non-DFT group. Conversely, the 2 groups were similar in age, left ventricular ejection fraction at baseline, functional class, and underlying cardiovascular disease. Results during a 12-month follow-up, 13 (31.0%) and 30 (37.5%) ventricular tachyarrhythmic episodes were recorded in the DFT and non-DFT groups, respectively (P = .472). Antitachycardia pacing (ATP) terminated most of episodes, reducing the need of defibrillation at 7.7% in the DFT group and 3.3% in the non-DFT group (P = .533). Similar percentages of inappropriate ATP interventions (7.1% vs 3.8%, P = .413) and shock deliveries (2.4% vs 5.0%, P = .659) were recorded between DFT and non-DFT groups.
CONCLUSIONS: At 1-year follow-up, the performance of DFT testing does not seem to add any significant efficacy advantage in patients undergoing ICD implantation. Prospective randomized trials and long-term follow-up are warranted to clarify whether routine DFT testing may be safely abandoned leading to a revision of current guidelines.
METHODS: We analyzed data obtained from follow-up visits of 122 consecutive patients who underwent ICD implantation at our institute from April 1996 to June 2008, with (n = 42) or without (n = 80) DFT testing. Patients in the DFT group were less likely to be men (83.3% vs 96.3%, P < .031) than those in the non-DFT group. Conversely, the 2 groups were similar in age, left ventricular ejection fraction at baseline, functional class, and underlying cardiovascular disease. Results during a 12-month follow-up, 13 (31.0%) and 30 (37.5%) ventricular tachyarrhythmic episodes were recorded in the DFT and non-DFT groups, respectively (P = .472). Antitachycardia pacing (ATP) terminated most of episodes, reducing the need of defibrillation at 7.7% in the DFT group and 3.3% in the non-DFT group (P = .533). Similar percentages of inappropriate ATP interventions (7.1% vs 3.8%, P = .413) and shock deliveries (2.4% vs 5.0%, P = .659) were recorded between DFT and non-DFT groups.
CONCLUSIONS: At 1-year follow-up, the performance of DFT testing does not seem to add any significant efficacy advantage in patients undergoing ICD implantation. Prospective randomized trials and long-term follow-up are warranted to clarify whether routine DFT testing may be safely abandoned leading to a revision of current guidelines.
Предметные рубрики MeSH
AgedCardiac Pacing, ArtificialCardiomyopathiesCohort StudiesDefibrillators, ImplantableEquipment SafetyFollow-Up StudiesHeart Function TestsHumansIntraoperative CareMaleMiddle AgedMultivariate AnalysisOdds RatioRegistriesRetrospective StudiesRisk AssessmentSensory ThresholdsStroke VolumeTreatment OutcomeUnnecessary ProceduresVentricular Fibrillation