MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2019.03.023

Медицинская библиотека · JACC. Cardiovascular interventions · 2019

Edge-to-Edge Mitral Valve Repair With Extended Clip Arms: Early Experience From a Multicenter Observational Study.

Praz Fabien, Braun Daniel, Unterhuber Matthias, Spirito Alessandro, Orban Mathias, Brugger Nicolas, Brinkmann Isabel, Spring Karin, Moschovitis Aris, Nabauer Michael, Blazek Stephan, Pilgrim Thomas, Thiele Holger, Lurz Philipp, Hausleiter Jörg, Windecker Stephan

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to investigate the technical success and efficacy of mitral valve edge-to-edge repair using extended clip arms.

BACKGROUND

A new iteration of the MitraClip system, the MitraClip XTR, was introduced in 2018 with the aim of addressing technical limitations observed with previous versions.

METHODS

Patients having received at least 1 new implant for the treatment of symptomatic mitral regurgitation (MR) were eligible for this study.

RESULTS

Among the 107 patients (mean age 76 ± 9 years, 69% men) included in this study, the etiology of MR was balanced, with one-half (n = 53 [50%]) classified as secondary and the remaining 54 patients having either primary (n = 40 [37%]) or mixed (n = 14 [13%]) disease. The mean number of devices implanted was 1.5 ± 0.6. Multiple device implantation was required in 46 patients (43%). Single-leaflet device attachment occurred in 4 patients and leaflet injury in 2 additional patients, requiring surgical conversion in 4 patients. Among the 102 patients discharged alive without mitral valve surgery, 95 (93%) had MR ≤2+ and 79 (77%) had MR ≤1+. The mean transmitral gradient increased from 1.9 ± 1.0 mm Hg at baseline to 3.5 ± 1.8 mm Hg at discharge (p < 0.001).

CONCLUSIONS

Technical success with the new mitral valve repair system with extended clip arm was achieved in 93% of the patients. MR ≤2+ was obtained in 95 patients (93%) and MR ≤1+ in 79 (77%). The main reasons for procedural failure were acute single-leaflet device attachment associated with leaflet damage or isolated leaflet injury and often required surgical correction.

Для цитирования: Praz Fabien, Braun Daniel, Unterhuber Matthias, Spirito Alessandro, Orban Mathias, Brugger Nicolas, Brinkmann Isabel, Spring Karin, Moschovitis Aris, Nabauer Michael, Blazek Stephan, Pilgrim Thomas, Thiele Holger, Lurz Philipp, Hausleiter Jörg, Windecker Stephan. Edge-to-Edge Mitral Valve Repair With Extended Clip Arms: Early Experience From a Multicenter Observational Study.. JACC. Cardiovascular interventions. 2019. DOI: 10.1016/j.jcin.2019.03.023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.