DOI 10.1002/jcla.24610

Медицинская библиотека · Journal of clinical laboratory analysis · 2022

A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators.

Findeisen Peter, Leis Marta, Bendig Garnet, Grimme Jill, Moser Elizabeth, Vogl Christian, Christenson Robert

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Vitamin D deficiency/insufficiency and toxicity are worldwide issues; thus, accurate diagnostic assays are required to measure vitamin D. We evaluated the performance of the new Elecsys® Vitamin D total III assay (Roche Diagnostics International Ltd).

METHODS

Repeatability and intermediate precision of the Elecsys Vitamin D total III assay (cobas e 601 analyzer) were evaluated at three sites using five human serum pools (HSPs) and two PreciControls (five-day model, one reagent lot [CLSI-EP05-A3]) and compared against prespecified acceptance criteria. A serum verification panel, with reference isotope dilution liquid chromatography tandem mass spectrometry (ID-LC-MS/MS) values, was used for comparator assay/concordance studies at two sites, assessed using unweighted Deming regression. Testing of serum vs. plasma on the Elecsys assay was conducted at one site using samples from healthy adults; assessed using Passing-Bablok regression.

RESULTS

Repeatability (HSP1 [16.8-18.4 ng/ml], SD 0.87-1.07; HSP5 [94.5-98.0 ng/ml], CV 1.58%-2.76%) and intermediate precision (HSP1, SD 1.14-1.77; HSP5, CV 2.00%-4.13%) met acceptance criteria across sites. Agreement was observed between the Elecsys assay and (i) the ID-LC-MS/MS verification panel (slope, 0.936-1.01; Pearson's r, 0.960-0.986) and (ii) comparator assays (slope, 0.921-1.15; Pearson's r, 0.958-0.982). The Elecsys assay correctly assigned the highest combined percentage of samples to deficient (100%) and insufficient (89.5%) vitamin D categories vs. comparator assays and demonstrated comparable performance in serum and plasma (y = 0.103 + 0.984x).

CONCLUSIONS

The Elecsys Vitamin D total III assay demonstrated good analytical performance and compared favorably with other assays, supporting its use in clinical practice.

Для цитирования: Findeisen Peter, Leis Marta, Bendig Garnet, Grimme Jill, Moser Elizabeth, Vogl Christian, Christenson Robert. A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators.. Journal of clinical laboratory analysis. 2022. DOI: 10.1002/jcla.24610.