Pathology of balloon-expandable transcatheter aortic valves.

Yahagi Kazuyuki, Ladich Elena, Kutys Robert, Mori Hiroyoshi, Svensson Lars G, Mack Michael J, Herrmann Howard C, Smith Craig R, Leon Martin B, Virmani Renu, Finn Aloke V

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The Placement of AoRtic TraNscathetER Valves trials (PARTNER) showed favorable safety and efficacy versus medical or surgical therapy in inoperable, high, and intermediate surgical risk patients with severe aortic stenosis. However, the biological responses to transcatheter aortic valves have not been well characterized.

OBJECTIVES

The aim of this study was to perform pathologic assessment of Edwards SAPIEN transcatheter aortic valves removed either at autopsy or surgically during the PARTNER I and II clinical trials.

METHODS

Explanted valves and frame were evaluated for pathologic responses including extent of thrombus, inflammation, neointima, and leaflet degeneration/calcification according to semiquantitative grading by implant duration (≤30 days; 31-90 days; >90 days).

RESULTS

A total of 22 cases (median age 82.0 years, 45% men) were included, with a duration of implantation that ranged from 0 to 1739 days (median duration 16.5 days [interquartile range, 2.8-68.3]). Valve thrombosis resulting in severe aortic stenosis was observed in one case. Moderate leaflet thrombus was seen in 14% of cases (n = 3) and all were asymptomatic. Calcification was seen in two valves: one with severe leaflet calcification had severe aortic stenosis requiring surgical replacement, while the other showed early calcification. Mild structural leaflet changes were exclusively seen in valve implants >90 days. Valve inflammation and thrombus formation was mild in majority of the cases.

CONCLUSIONS

Overall, our study demonstrates moderate thrombus formation in 14% and calcification in only 2 valves, ≥4 years duration. In this short-duration study, acceptable durability and biocompatibility of the Edwards SAPIEN transcatheter valve system was demonstrated; however, further studies are required to confirm the significance and application of our findings.

Для цитирования: Yahagi Kazuyuki, Ladich Elena, Kutys Robert, Mori Hiroyoshi, Svensson Lars G, Mack Michael J, Herrmann Howard C, Smith Craig R, Leon Martin B, Virmani Renu, Finn Aloke V. Pathology of balloon-expandable transcatheter aortic valves.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2017. DOI: 10.1002/ccd.27160.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.