Temporary atrial electrode for the treatment of supraventricular tachycardia after cardiac operations.

Mehmanesh H, Lange R, Hagl S

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Supraventricular tachycardia is a common postoperative complication early after cardiac operations. A temporary atrial patch electrode for low-energy atrial defibrillation was developed in 1992 and subsequently tested.

METHODS

The electrode first was tested and removed intraoperatively during open heart operations in 10 patients (phase I). After the intraoperative testing, the temporary atrial patch electrode was implanted in 20 patients for postoperative termination of spontaneous episodes of supraventricular tachycardia (phase II). When supraventricular tachycardia occurred, biphasic shocks (1.2 to 5 J) were applied and the atrial defibrillation thresholds were measured.

RESULTS

In phase I, the mean intraoperative atrial defibrillation threshold was 1.6 +/- 1.4 J, with a mean shock impedance of 64 +/- 7.3 omega. In phase II, 6 of 20 patients (30%) had 7 episodes of atrial fibrillation (n = 6) and atrial flutter (n = 1) after operation. In 5 patients, the supraventricular tachycardia could be converted to a sinus (n = 5) or normofrequent atrioventricular rhythm (n = 1). The mean postoperative defibrillation threshold was 2.7 +/- 2.1 J, with a mean shock impedance of 50.2 +/- 6.8 omega.

CONCLUSIONS

The temporary atrial patch electrode allows low-energy defibrillation of episodes of atrial fibrillation. It may serve as an alternative therapeutic option for the treatment of supraventricular tachycardia.

Для цитирования: Mehmanesh H, Lange R, Hagl S. Temporary atrial electrode for the treatment of supraventricular tachycardia after cardiac operations.. The Annals of thoracic surgery. 1998. DOI: 10.1016/s0003-4975(97)01350-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.