MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00405-022-07265-2

Медицинская библиотека · European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2022

Evaluation of the Bonebridge BCI 602 active bone conductive implant in adults: efficacy and stability of audiological, surgical, and functional outcomes.

Cywka Katarzyna B, Skarzynski Piotr H, Krol Bartlomiej, Hatzopoulos Stavros, Skarzynski Henryk

СтатьяПеревод в очереди

Аннотация

PURPOSE: (1) To assess the effectiveness and safety of a bone-conduction implant, the Bonebridge BCI 602, in adults with conductive or mixed hearing loss. (2) To investigate whether the Bonebridge BCI 602 is at least as effective as the Bonebridge BCI 601 in such patients.

METHODS

The study group included 42 adults who had either conductive or mixed hearing loss. All patients underwent Bonebridge BCI 602 implant surgery. Before and after implantation, pure-tone audiometry, speech recognition tests (in quiet and noise), and free-field audiometry were performed. Word recognition scores were evaluated using the Polish Monosyllabic Word Test. Speech reception thresholds in noise were assessed using the Polish Sentence Matrix Test. Subjective assessment of benefits was done using the APHAB (Abbreviated Profile of Hearing Aid Benefit) questionnaire.

RESULTS

The APHAB questionnaire showed that difficulties in hearing decreased after BCI 602 implantation. Both word recognition in quiet and speech reception threshold in noise were significantly better after BCI 602 implantation and remained stable for at least 12 months. A significant advantage of the device is a reduced time for surgery while maintaining safety. In this study, the mean time for BCI 602 implantation was 28.3 min ± 9.4.

CONCLUSIONS

The second-generation Bonebridge BCI 602 implant is an effective hearing rehabilitation device for patients with conductive or mixed hearing loss. Patient satisfaction and audiological results confirm its efficacy and safety. Its new shape and dimensions allow it to be used in patients previously excluded due to insufficient or difficult anatomical conditions. The new BCI 602 implant is as effective as its predecessor, the BCI 601.

Для цитирования: Cywka Katarzyna B, Skarzynski Piotr H, Krol Bartlomiej, Hatzopoulos Stavros, Skarzynski Henryk. Evaluation of the Bonebridge BCI 602 active bone conductive implant in adults: efficacy and stability of audiological, surgical, and functional outcomes.. European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery. 2022. DOI: 10.1007/s00405-022-07265-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.